首页> 外文期刊>Cardiology and therapy. >Inappropriate ICD Discharge Related to Electrical Muscle Stimulation in Chiropractic Therapy: A Case Report
【24h】

Inappropriate ICD Discharge Related to Electrical Muscle Stimulation in Chiropractic Therapy: A Case Report

机译:整脊疗法中与电刺激肌肉相关的ICD放电不当:一例报告

获取原文
           

摘要

Implantable cardioverter defibrillators (ICDs) have been shown to reduce mortality in high-risk patients. Despite several advances in programming ICDs, inappropriate shocks persist and continue to be psychologically and physically disturbing. External electromagnetic interference from electrocautery, welding, acupunctures, low-output transcutaneous electric nerve stimulators, and electronic muscle stimulators may result in inappropriate ICD sensing and shock. We present a 63-year-old female who presented to the emergency department after an ICD shock while undergoing electronic muscle stimulation in chiropractic treatment, during which light electrical pulses were sent through skin electrodes. Our case highlights the importance of recognizing methods used by alternative medicine doctors, including electrical muscle stimulation that may cause electromagnetic interference and inappropriate ICD discharge and therefore, a higher overall mortality risk.
机译:植入式心脏复律除颤器(ICD)已被证明可以降低高危患者的死亡率。尽管在对ICD进行编程方面取得了一些进步,但不适当的冲击仍然存在,并且继续在心理和身体上造成干扰。电灼,焊接,针灸,输出功率低的经皮电神经刺激器和电子肌肉刺激器产生的外部电磁干扰可能会导致不适当的ICD感应和电击。我们介绍了一位63岁的女性,该患者在ICD休克后在急诊治疗中接受电子肌肉刺激时出现在急诊室,其间通过皮肤电极发送光电脉冲。我们的案例凸显了识别替代医学医生使用的方法的重要性,包括可能会引起电磁干扰和不适当的ICD放电的肌肉电刺激,因此总体死亡率较高。