首页> 外文期刊>Journal of interventional cardiac electrophysiology: an international journal of arrhythmias and pacing >Inappropriate implantable cardioverter defibrillator shock from a transcutaneous muscle stimulation device therapy.
【24h】

Inappropriate implantable cardioverter defibrillator shock from a transcutaneous muscle stimulation device therapy.

机译:从经皮肌刺激装置治疗中不恰当的植入式心脏除颤器冲击。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Inappropriate shock from implantable cardioverter defibrillator (ICD) may result from external electromagnetic interference (EMI), especially for unipolar ventricle sensing. Previous case reports and small in-vitro safety study suggested that endocardial bipolar lead system may be immune from EMI resulting from transcutaneous electrical neuromuscle stimulation (TENS) therapy. This report presents an unusual case of inappropriate discharge in a patient with ICD of endocardial bipolar lead system, receiving TENS from a commercially available device.
机译:来自植入式心脏除颤器(ICD)的不恰当震荡可能是由外部电磁干扰(EMI)产生的,特别是对于单极性心室感测。 以前的病例报告和小的体外安全性研究表明,内心膜膜双极铅系统可以免受经皮电子神经血型刺激(TENS)治疗产生的EMI免疫。 本报告提出了一个不寻常的案例,在具有ICD的Condocardial双极铅系统中的患者中的患者,从市售的设备接收数十。