Methods of measuring the iridocorneal angle in tomographic images of the anterior segment of the eye

机译:测量眼前节断层图像中虹膜角膜角的方法

获取原文
           

摘要

Introduction This paper presents the problem of automatic measurement of the iridocorneal angle in tomographic images of the anterior segment of the eye. It includes the results of the comparison of well-known methods for measuring the iridocorneal angle with new methods, proposed in this paper. All these methods concern tomographic image analysis and processing. Material and method In total, approximately 100’000 tomographic images (from about 6’000 patients) were analysed. They were obtained using two devices: SOCT Copernicus (Optopol Tech. SA, Zawiercie, Poland) and Visante OCT (Carl Zeiss Meditec, Inc, Dublin, California, USA). The patients, aged 12 to 78?years with varying degrees of the iridocorneal angle pathology, were from the region of Silesia, Poland. The images were in DICOM or RAW formats and analysed in the software developed by the authors for the purposes of this study. Results The results indicate that the measurement method proposed by the authors, which is based on the calculation of the minimum distance between the iris and the cornea in the adopted area, is the most accurate. For this method sensitivity was 0.88, specificity 0.89 and the area under the Receiver Operating Characteristic curve (AUC) was 0.88. The other known methods for measuring the iridocorneal angle gave worse results, that is, for example, for the measurement of the distance between the iris and the cornea AUC?=?0.87, sensitivity?=?0.86 and specificity?=?0.71. For another well-known method of measuring the iridocorneal angle AUC?=?0.77, sensitivity?=?0.82 and specificity?=?0.61. Conclusions The study proved that the proposed method of measuring the minimum distance between the iris and the cornea within the adopted area is the most effective in the classification of the iridocorneal angle in patients with a high degree of pathology of all the compared measurement methods based on tomographic images. However, it requires fully automated measurement.
机译:引言本文提出了在眼前节的断层图像中自动测量虹膜角膜角的问题。它包括了本文提出的用新方法测量虹膜角膜角的著名方法的比较结果。所有这些方法都涉及断层图像分析和处理。材料和方法总共分析了约100'000幅断层图像(来自约6000例患者)。它们是使用两种设备获得的:SOCT Copernicus(波兰Zawiercie的Opopol Tech。SA)和Visante OCT(美国加利福尼亚州都柏林的卡尔蔡司Meditec,Inc)。这些患者年龄在12至78岁之间,具有不同程度的虹膜角膜角病变,来自波兰西里西亚地区。图像采用DICOM或RAW格式,并在作者为本研究目的开发的软件中进行了分析。结果结果表明,作者提出的测量方法是基于对所采用区域中虹膜和角膜之间的最小距离的计算得出的,是最准确的。对于此方法,灵敏度为0.88,特异性为0.89,接收器工作特征曲线(AUC)下的面积为0.88。测量虹膜角膜角的其他已知方法给出了较差的结果,例如,用于测量虹膜和角膜之间的距离的AUCα= 0.87,灵敏度α= 0.86和特异性α= 0.71。对于另一种测量虹膜角膜角AUCα=α0.77的方法,灵敏度α=α0.82和特异性α=α0.61。结论研究证明,在所有基于此方法的比较测量方法中,对于高度病理的患者,所提出的测量虹膜与角膜之间的最小距离的方法在对虹膜角膜角分类中最有效。断层图像。但是,它需要全自动测量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号