首页> 外文期刊>Biosensors Journal >Anti-Prostate Specific Antigen (Anti-PSA) Modified Interdigitated Microelectrode-Based Impedimetric Biosensor for PSA Detection
【24h】

Anti-Prostate Specific Antigen (Anti-PSA) Modified Interdigitated Microelectrode-Based Impedimetric Biosensor for PSA Detection

机译:抗前列腺特异性抗原(Anti-PSA)修饰的基于叉指微电极的阻抗生物传感器,用于PSA检测

获取原文
           

摘要

Interdigitated microelectrodes (IDEs) functionalized with a cysteamine (Cys) self-assembled monolayer (SAM) were used to fabricate a sensitive, disposable, impedimetric biosensor for prostate specific antigen (PSA) detection. PSA specific monoclonal antibody (anti-PSA) was covalently immobilized on cysteamine modified gold surface of IDEs, using N-ethyl-N’-(3-dimethylaminopropyl) carbodiimide and N-hydroxysuccinimide (EDC/NHS) chemistry. Bovine serum albumin (BSA) was used as a blocker to prevent nonspecific adsorption on electrode surface. BSA-anti-PSA/Cys/IDE electrodes were exposed to solutions with different PSA concentrations and their response was measured using a label-free electrochemical impedance (EIS) technique. The impedance changes revealed close a correlation with PSA concentrations in the measurement range of 1 pg.mL?1–100 ng.mL?1, with sensitivity of 0.444 mL.g?1.
机译:用半胱胺(Cys)自组装单层(SAM)功能化的叉指式微电极(IDE)用于制造灵敏的,一次性的,阻抗测定的生物传感器,用于检测前列腺特异性抗原(PSA)。使用N-乙基-N’-(3-二甲基氨基丙基)碳二亚胺和N-羟基琥珀酰亚胺(EDC / NHS)化学方法,将PSA特异性单克隆抗体(anti-PSA)共价固定在IDE的半胱胺修饰的金表面上。牛血清白蛋白(BSA)被用作阻滞剂,以防止电极表面发生非特异性吸附。将BSA-anti-PSA / Cys / IDE电极暴露于具有不同PSA浓度的溶液中,并使用无标记电化学阻抗(EIS)技术测量其响应。阻抗变化表明与PSA浓度在1 pg.mL?1–100 ng.mL?1的测量范围内密切相关,灵敏度为0.444 mL.g?1。