LC−UV Analysis of N-{3-(2,4-Dioxo-1,4-dihydro-2H-quinazolin-3-yl)propyl}-N-[4-{3-(2,4-dioxo-1,4-dihydro-2H-quinazolin-3-yl)propyl-amino}butyl]acetamide (SP-8203) in Rat Plasma, Urine, and Gastrointestinal Tract Samples

机译:N- {3-(2,4-Dioxo-1,4-dihydro-2H-quinazolin-3-yl)propyl} -N- [4- {3-(2,4-dioxo-1)的LC-UV分析,血浆,尿液和胃肠道样品中的,4-二氢-2H-喹唑啉-3-基)丙基-氨基}丁基]乙酰胺(SP-8203)

获取原文
获取原文并翻译 | 示例

摘要

A simple, rapid, and reproducible reversed-phase LC method with UV detection at 215 nm has been developed for analysis of SP-8203 in rat samples. A C18 column was used with 3,000:1,050 (v/v) 0.01 m K2HPO4 buffer (pH 3)–acetonitrile as mobile phase at a flow rate of 1.7 mL min−1 at 50 °C. Samples were extracted with dichloromethane containing ondansetron (internal standard). Detection limits for SP-8203 in plasma, urine, and gastrointestinal tract samples were 0.05, 0.5, and 10 μg mL−1, respectively. The method was suitable for pharmacokinetic study of SP-8203 in rats after intravenous administration. Keywords Column liquid chromatography - SP-8203 - Rat plasma, urine, and gastrointestinal tract
机译:已经开发了一种简单,快速且可重现的反相LC方法,该方法在215 nm处具有UV检测,可用于分析大鼠样品中的SP-8203。 AC 18 柱使用3,000:1,050(v / v)0.01 m K 2 HPO 4 缓冲液(pH 3)-乙腈作为流动介质相在50°C时的流速为1.7 mL min -1 。样品用含有恩丹西酮(内标)的二氯甲烷萃取。血浆,尿液和胃肠道样品中SP-8203的检出限分别为0.05、0.5和10μgmL -1 。该方法适用于大鼠静脉给药后SP-8203的药代动力学研究。柱液相色谱-SP-8203-大鼠血浆,尿液和胃肠道

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号