首页> 外文期刊>Australasian physical & engineering sciences in medicine >Quantitative analysis of effects of the grid specifications on the quality of digital radiography images
【24h】

Quantitative analysis of effects of the grid specifications on the quality of digital radiography images

机译:网格规范对数码造影图像质量影响的定量分析

获取原文
获取原文并翻译 | 示例
           

摘要

A grid is one of the key components of a digital radiography (DR) system because it removes scattered radiation, which arises when X-rays penetrate an object and improves diagnostic accuracy by enhancing image quality. With the widespread use of DR systems, demand for grids with high precision has simultaneously increased. Because unsuitable grids may decrease image quality and lead to misdiagnosis, using optimised grids for DR systems is critical. In this study, we aimed to analyse the quality of X-ray images acquired using grids with different specifications and proposed standardised criteria for grid use on the basis of our results. We measured modulation transfer function (MTF), normalised noise power spectrum (NNPS) and detective quantum efficiency (DQE) using grids with different ratios (10:1, 12:1 and 15:1) with or without implementing poly methyl methacrylate (PMMA) phantoms (0-20cm). Pixel pitch of the detector used in this experiment was 143m. Based on this, a grid with a line frequency of 85 line pairs/cm was selected to prevent distortion caused by implementing unoptimised grids. As a result, the NNPS was found to increase when using the grid, and the difference in MTF and DQE was only measured when the scattered X-ray was generated by stacking the PMMA phantom. However, grids showed a positive effect MTF and DQE when the PMMA phantom was implemented. Specifically, MTF and DQE improved with increase in grid ratio. Thus, it is desirable to use a high-ratio grid to improve image quality.
机译:网格是数字射线照相(DR)系统的关键组件之一,因为它去除散射辐射,当X射线穿透对象时出现并通过提高图像质量来提高诊断精度。随着DR系统的广泛使用,对具有高精度的网格的需求同时增加。因为不合适的网格可能会降低图像质量并导致误诊,使用用于DR系统的优化网格至关重要。在这项研究中,我们旨在分析使用不同规格的网格获取的X射线图像的质量,并在我们的结果的基础上提出了网格使用的标准化标准。我们使用具有不同比率(10:1,12:1和15:1)的网格测量调制传递函数(MTF),归一化噪声功率谱(NNP)和侦探量子效率(DQE),或者不具有聚甲基丙烯酸甲酯(PMMA )幽灵(0-20cm)。本实验中使用的检测器的像素间距为143M。基于此,选择具有85线对/ cm的线频率的网格,以防止通过实现未优化的网格引起的失真。结果,发现NNPS在使用网格时增加,并且仅在通过堆叠PMMA幻像产生散射X射线时测量MTF和DQE的差异。然而,当实施PMMA幻影时,网格显示出积极的影响MTF和DQE。具体而言,MTF和DQE随着网格比的增加而得到改善。因此,希望使用高比率网格来改善图像质量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号