Commissioning of a well type chamber for HDR and LDR brachytherapy applications: a review of methodology and outcomes

机译:用于HDR和LDR近距离放射治疗应用的井式腔室的调试:方法和结果的回顾

获取原文
获取原文并翻译 | 示例
           

摘要

For safe and accurate dose delivery in brachytherapy, associated equipment is subject to commissioning and ongoing quality assurance (QA). Many centres depend on the use of a well-type chamber ('well chamber') for performing brachytherapy dosimetry. Documentation of well chamber commissioning is scarce despite the important role the chamber plays in the whole brachytherapy QA process. An extensive and structured commissioning of the HDR 1000 plus well chamber (Standard Imaging Inc, Middleton WI) for HDR and LDR dosimetry was undertaken at Sir Charles Gairdner Hospital. The methodology and outcomes of this commissioning is documented and presented as a guideline to others involved in brachytherapy. The commissioning tests described include mechanical integrity, leakage current, directional dependence, response, length of uniform response, the influence of insert holders, ion collection efficiency, polarity effect, accuracy of measured air kerma strength (S (K) ) or reference air kerma rate (K (R) ) and baseline setting (for ongoing constancy checks). For the HDR 1000 plus well chamber, some of the insert holders modify the response curve. The measured sweet length was 2.5 cm which is within 0.5 % of that specified by the manufacturer. Correction for polarity was negligible (0.9999) and ion recombination was small (0.9994). Directional dependence was small (less than 0.2 %) and leakage current was negligible. The measured K (R) for Ir-192 agreed within 0.11 % compared with a second well chamber of similar model and was within 0.5 % of that determined via a free-in-air measurement method. Routine constancy checks over a year agreed with the baseline within 0.4 %.
机译:为了在近距离放射治疗中安全准确地输送剂量,相关设备必须进行调试和持续的质量保证(QA)。许多中心都依赖使用井式腔室(“ well chamber”)进行近距离放射治疗剂量测定。尽管腔室在整个近距离放射治疗质量保证过程中发挥着重要作用,但缺乏关于腔室调试的文件。 Charles Gairdner爵士医院对HDR 1000 plus井室(标准成像公司,威斯康星州Middleton)进行了广泛而结构化的调试,以进行HDR和LDR剂量测定。此调试的方法和结果已记录在案,并作为其他近距离放射治疗的指南。描述的调试测试包括机械完整性,泄漏电流,方向依赖性,响应,均匀响应的长度,刀片支架的影响,离子收集效率,极性效应,测量的空气比释动能强度(S(K))或参考空气比释动能的准确性速率(K(R))和基线设置(用于持续的稳定性检查)。对于HDR 1000 plus井室,某些插入物固定器会修改响应曲线。测得的甜味长度为2.5cm,在制造商指定的甜味长度的0.5%以内。极性校正可忽略不计(0.9999),离子重组很小(0.9994)。方向依赖性小(小于0.2%),漏电流可忽略不计。与类似模型的第二个井室相比,Ir-192的测得K(R)在0.11%之内,并且是通过自由空气测量方法测得的K(R)的0.5%之内。一年中进行的常规稳定性检查在0.4%的范围内与基线相符。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号