Some well-type ionization chambers, present a very small sweet spot that are sufficient for small HDR sources. However, if a longer HDR source or LDR wires are calibrated, the positional uncertainty increases and an approximated correction factor must be applied, resulting in an increased uncertainty. One of the ways to avoid this problem would be to flatten the well chamber response by increasing its sweet spot region. This work uses the Monte Carlo code PENELOPE to simulate the response of a well-type chamber HDR-1000, with its original insert, by using an HDR 192Ir source and proposes a new insert design that increases its flatness region from 1.0 cm to approximately 4.0 cm (+/- 2.0 cm about the peak response).
机译:一些井型电离室呈现出非常小的甜点,足以满足小型HDR源的需要。但是,如果校准了更长的HDR源或LDR导线,则位置不确定性会增加,并且必须应用近似的校正系数,从而导致不确定性增加。避免此问题的方法之一是通过增加其最佳点区域来使井室响应变平坦。这项工作使用Monte Carlo代码PENELOPE通过使用HDR 192Ir源来模拟带有原始插入物的井式腔室HDR-1000的响应,并提出了一种新的插入物设计,该设计将其平坦度区域从1.0 cm增加到了大约4.0厘米(关于峰值响应+/- 2.0厘米)。