首页> 外文期刊>Artificial life and robotics >Position detection method for microwell chip on microscope stage by photon multiplier tube
【24h】

Position detection method for microwell chip on microscope stage by photon multiplier tube

机译:光子倍增管在显微镜载物台上微孔芯片的位置检测方法

获取原文
获取原文并翻译 | 示例
           

摘要

In this paper, we propose a position detection method for a microwell with manageable volume (MMV) chip on the stage of the inverted microscope. When the manipulation robot arm puts an MMV chip on the microscope stage, there is alignment error up to 500 μm. This error affects measurement accuracy of fluorescence from solution filled in a tiny hall of the MMV chip. Usually, CCD image processing is used for detecting a position. However, appending a CCD camera to our microscope increases the complexity of the optical system. Therefore, we used a photon multiplier tube which was already equipped for measuring fluorescence. In the result, the position of the MMV chips was detected by putting a pilot LED directly above the objective lens. The detection error was less than 10 μm. Although this method was devised for our system, we considered that it was available in other micro optics systems.
机译:在本文中,我们提出了一种在倒置显微镜载物台上对具有可控体积(MMV)芯片的微孔进行位置检测的方法。当操纵机器人手臂将MMV芯片放在显微镜载物台上时,会出现高达500μm的对准误差。该误差会影响来自MMV芯片微小大厅中填充溶液的荧光的测量精度。通常,CCD图像处理用于检测位置。但是,将CCD相机附加到我们的显微镜会增加光学系统的复杂性。因此,我们使用了已经配备了用于测量荧光的光子倍增管。结果,通过在物镜上方直接放置一个指示灯来检测MMV芯片的位置。检测误差小于10μm。尽管此方法是为我们的系统设计的,但我们认为其他微光学系统也可以使用此方法。