Introduction: Improving the longevity and reliability of cemented total knee arthroplasty (TKA) remains a major step to achieve. It is still unclear, whether a cemented tibial stem reduces micromotion of the tibial tray and produces therefore a better initial stability or not. The higher conformity of rotating platform design and the possible rotary forces to the tibial platform may produce higher micromotion when the tibial stem remains cementless (hybrid fixation). Materials and methods: An in vitro study was performed using the PFC® mobile bearing tibial tray (DePuy®, Warswa, IN, USA) to test the hypothesis that the addition of cement surrounding the tibial stem reduces micromotion of the tibial tray in cemented TKA with mobile bearing design. Ten tibial trays with mobile design were implanted in sawbones with a 3-mm cement mantle beneath the baseplate of the tibial tray and with or without the cemented stem. Tibial trays were loaded additionally in the ventral, lateral, medial and posterior positions with 2,500 N using the Zwick Z010® instrumentation and HBM pick up Hottinger Baldwin®. Results: In this study, a significant increased mean maximum liftoff was found when only cementing the tibial baseplate (hybrid fixation), compared to the fully cemented tibial tray (P<0.02). Conclusion: In conclusion, the stem of mobile bearing tibial components should be cemented to provide increased micromotion and earlier loosening.
机译:简介:提高骨水泥全膝关节置换术(TKA)的寿命和可靠性仍然是要实现的主要步骤。胶合的胫骨柄是否会降低胫骨托的微动并因此产生更好的初始稳定性仍不清楚。当胫骨柄保持非骨水泥状态(混合固定)时,旋转平台设计的较高一致性以及对胫骨平台可能的旋转力可能会产生较高的微运动。材料和方法:使用PFC®可移动的胫骨托(DePuy®,美国印第安纳州华沙)进行了一项体外研究,以检验以下假设:胫骨柄周围水泥的添加减少了骨水泥型TKA中胫骨托的微动,采用活动轴承设计。将十个可移动设计的胫骨托植入胫骨托,在胫骨托底板下方有一个3 mm的水泥套,并带有或不带有骨干。使用ZwickZ010®仪器和HBM拾取器HottingerBaldwin®,在胫骨托的腹侧,外侧,内侧和后部分别加装2500N。结果:在这项研究中,与完全骨水泥化的胫骨托相比,仅骨水泥化胫骨底板(混合固定)时,平均最大抬高明显增加(P <0.02)。结论:总之,应将活动轴承胫骨组件的茎骨粘合起来,以提供增加的微动并更早地松动。

著录项

  • 来源
    《Archives of Orthopaedic and Trauma Surgery》 |2006年第1期|45-48|共4页
  • 作者单位

    Department of Orthopaedic Surgery University of Regensburg Kaiser-Karl-V. Allee 3 93077 Bad Abbach Germany;

    Department of Orthopaedic Surgery University of Regensburg Kaiser-Karl-V. Allee 3 93077 Bad Abbach Germany;

    Department of Orthopaedic Surgery University of Regensburg Kaiser-Karl-V. Allee 3 93077 Bad Abbach Germany;

    Department of Orthopaedic Surgery University of Regensburg Kaiser-Karl-V. Allee 3 93077 Bad Abbach Germany;

    Department of Orthopaedics Amtssygehus Aalborg Sygehus Aarhus Universitetshospital Hobrovej Aalborg Denmark;

    Laboratory for Biomechanics and Experimental Orthopaedics Ludwigs-Maximilians-University Munich Germany;

    Department of Orthopaedic Surgery University of Regensburg Kaiser-Karl-V. Allee 3 93077 Bad Abbach Germany;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Rotating platform; Total knee arthroplasty (TKA); Cementing technique; Micromotion;