A new type of ISFET-based differential pCO(2) sensor system is devised by employing a low-resistance gas-permeable membrane, Both the pCO(2)-FET and the reference FET (REFET) are made by coating their gate surfaces with an RTV silicone rubber film doped with valinomycin. The pCO(2) FET is covered with a hydrogel-based recipient layer inside the gas-permeable RTV membrane, but the REFET does not. include the hydrogel layer, Because of low resistance of the gas-permeable membrane, the reference electrode can be placed outside as a common reference for both FETs. Output signals from the pCO(2)-FET and the REFET are measured against the common reference, and the difference of these two signals is measured through a differential amplifier; therefore, calibration drifts of the two identical base ISFETs, caused by their thermal sensitivity, for example, cancel out It is also shown that the ion response or the potential drifts caused by both parts of the RTV membranes cancel out exactly the same way as in the case of the base ISFETs.
机译:通过使用低阻透气膜设计出一种新型的基于ISFET的差分pCO(2)传感器系统,pCO(2)-FET和参考FET(REFET)均通过在其栅极表面涂上一层掺杂了缬氨霉素的RTV硅橡胶薄膜。 pCO(2)FET在透气RTV膜内部覆盖有基于水凝胶的受体层,但REFET并未覆盖。由于透气膜的电阻低,参比电极可以放置在外部,作为两个FET的通用参比。来自pCO(2)-FET和REFET的输出信号是针对公共基准进行测量的,这两个信号的差值是通过差分放大器测量的。因此,两个相同的基本ISFET的校准漂移(例如,由其热敏性引起的)被抵消了。还显示出,由RTV膜的两个部分引起的离子响应或电位漂移的抵消方式与基本ISFET的情况。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号