首页> 美国卫生研究院文献>Wiley-Blackwell Online Open >Anatomy‐based patient‐specific VMAT QA using EPID or MLC log files
【2h】

Anatomy‐based patient‐specific VMAT QA using EPID or MLC log files

机译:使用EPID或MLC日志文件的基于解剖的针对患者的VMAT QA

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

In this project, we investigated the use of an electronic portal imaging device (EPID), together with the treatment planning system (TPS) and MLC log files, to determine the delivered doses to the patient and evaluate the agreement between the treatment plan and the delivered dose distribution. The QA analysis results are presented for 15 VMAT patients using the EPID measurements, the ScandiDos Delta4 dosimeter, and the beam fluence calculated from the multileaf collimator (MLC) log file. EPID fluence images were acquired in continuous acquisition mode for each of the patients and they were processed through an in‐house MATLAB program to create an opening density matrix (ODM), which was used as the input fluence for the dose calculation in the TPS (Pinnacle3). The EPID used in this study was the aSi1000 Varian on a Novalis TX linac equipped with high‐definition MLC. The actual MLC positions and gantry angles were retrieved from the MLC log files and the data were used to calculate the delivered dose distributions in Pinnacle. The resulting dose distributions were then compared against the corresponding planned dose distributions using the 3D gamma index with 3mm/3% passing criteria. The ScandiDos Delta4 phantom was also used to measure a 2D dose distribution for all the 15 patients and a 2D gamma was calculated for each patient using the Delta4 software. The average 3D gamma using the EPID images was 96.1%±2.2%. The average 3D gamma using the log files was 98.7%±0.5%. The average 2D gamma from the Delta4 was 98.1%±2.1%. Our results indicate that the use of the EPID, combined with MLC log files and a TPS, is a viable method for QA of VMAT plans.PACS numbers: 87.55.Qr
机译:在此项目中,我们调查了电子门禁成像设备(EPID)以及治疗计划系统(TPS)和MLC日志文件的使用,以确定向患者提供的剂量,并评估治疗计划与治疗方案之间的一致性。输送剂量分布。使用EPID测量,ScandiDos Delta 4 剂量计以及从多叶准直仪(MLC)日志文件计算出的光束通量,对15名VMAT患者进行了QA分析结果。以连续采集模式为每位患者采集EPID能量密度图像,并通过内部MATLAB程序对其进行处理,以创建开口密度矩阵(ODM),将其用作TPS中剂量计算的输入能量密度(尖峰 3 )。本研究中使用的EPID是配备高分辨率MLC的Novalis TX直线加速器上的aSi1000 Varian。从MLC日志文件中检索了实际的MLC位置和龙门架角度,并使用该数据计算了Pinnacle中的剂量分布。然后使用具有 3 mm / 3 通过标准。 ScandiDos Delta 4 体模还用于测量所有15位患者的2D剂量分布,并使用Delta 4 软件为每位患者计算2Dγ。使用EPID图像的平均3D伽玛值为 96.1 ± 2.2 。使用日志文件的平均3D伽玛值为 98.7 ± 0.5 。来自Delta 4 的平均2D伽玛为 98.1 ± 2.1 。我们的结果表明,结合使用EPID,MLC日志文件和TPS,是对VMAT计划进行质量检查的可行方法.PACS编号:87.55.Qr

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号