To enhance ambulation and facilitate hospital discharge of total artificial heart (TAH)–supported patients, we adapted a mobile ventricular assistance device (VAD) driver (Excor) for TAH use and report on the performance of Excor-driven TAH patients discharged home. Ten patients stabilized on a TAH, driven by the CSS (“Circulatory Support System”), were progressively switched over to the Excor in hospital over 14 days as a pilot, with daily hemodynamics and laboratory parameters measured. Twenty-two stable TAH patients were subsequently placed on the Excor, trained, and discharged home. Clinical and hemodynamic parameters were followed. All pilot study patients were clinically stable on the Excor, with no decrease in TAH output noted (6.3 + 0.3 L/min [day 1] vs. 5.8 + 0.2 L/min [day 14], p = 0.174), with a trend suggesting improvement of both hepatic and renal function. Twenty-two TAH patients were subsequently successfully discharged home on the portable driver and were supported out of hospital for up to 598 days (range, 2–598; mean = 179 ± 140 days), remaining ambulatory, New York Heart Association (NYHA) class I or II, and free of readmission for 88.5% of the time of support. TAH patients may be effectively and safely supported by a mobile drive system. As such, the utility of the TAH may be extended to support patients beyond the hospital, at home, with overall ambulatory freedom.
机译:为增强行走能力并促进人工心脏(TAH)支持的患者出院,我们为TAH使用了可移动的心室辅助设备(VAD)驱动器(Excor),并报告了由Excor驱动的TAH患者在家中的表现。由CSS(“循环支持系统”)驱动的10名靠TAH稳定的患者在14天的时间内被逐步转移到医院的Excor,进行每日血流动力学和实验室参数测量。随后,有22名稳定的TAH患者被放到Excor上接受训练,然后出院回家。遵循临床和血液动力学参数。所有先导研究患者均在Excor上临床稳定,未观察到TAH输出降低(6.3 + 0.3 L / min [第1天]与5.8 + 0.2 L / min [第14天],p = 0.174),并且趋势提示肝脏和肾脏功能均得到改善。随后,有22名TAH患者通过便携式驾驶员成功出院,并得到了长达598天的住院支持(范围为2–598;平均= 179±140天),其余病友仍在纽约心脏协会(NYHA) I级或II级,并在88.5%的支持时间内免收再次入学。 TAH患者可以通过移动驱动系统得到有效而安全的支持。这样,TAH的效用可以扩展到以整体的门诊自由度来支持患者在家中医院以外。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号