The ABL 500 blood gas analyser from Radiometer has cordless electrodes and does not use a humidifier for calibrating gases. During the evaluation of the analytical performance of this instrument, the problem of p02 accuracy was approached by comparing the values obtained with two kinds of tonometry (film and bubble). An acceptable level of imprecision was demonstrated for all measured parameters. For within-run precision, with tonometry, coefficients of variation (CV) were ≆0.37% for pO2 and ≆0.52% for pCO2. A CV of 1.76% was found for day-to-day precision for both p02 and pCO2. In the linearity study, with both tonometry methods, and in the inter-instrument comparisons (the ABL was compared with the Ciba Corning 178), pO2 values obtained on the ABL 500 exhibited a slight overestimation above 150 mmHg (2.2-3.4% at 600 mmHg). This minor discrepancy is discussed with reference to the new design of the pO2 electrode, the algorithm for pO2 correction and the tonometry procedure. The results reported in this paper stress the importance of pO2 accuracy assessment for the evaluation of blood gas analysers.
机译:辐射计公司的ABL 500血气分析仪具有无线电极,不使用加湿器来校准气体。在评估该仪器的分析性能时,通过比较两种眼压计(膜和气泡)获得的值来解决p02精度问题。对于所有测量参数,证明了可接受的不精确度。对于运行内精度,使用眼压计,pO2的变异系数(CV)为≆0.37%,pCO2的变异系数(CV)为≆0.52%。发现p02和pCO2的日常精度CV为1.76%。在线性研究中,通过两种眼压测量方法以及仪器间的比较(将ABL与Ciba Corning 178进行了比较),在ABL 500上获得的pO2值均略高于150 mmHg(在600时为2.2-3.4%)毫米汞柱)。参照pO2电极的新设计,pO2校正算法和眼压测量程序讨论了这种微小差异。本文报道的结果强调了pO2准确性评估对评估血气分析仪的重要性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号