首页> 美国卫生研究院文献>International Orthopaedics >A proximal fixed anatomic femoral stem reduces stress shielding
【2h】

A proximal fixed anatomic femoral stem reduces stress shielding

机译:股骨近端解剖解剖固定柄减少了应力屏蔽

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

In 24 patients with total hip replacement using a short anatomic femoral stem, bone mineral density (BMD) was measured after a 7-year follow-up using dual-energy X-ray absorptiometry. The contralateral side was used as a control. The BMD on the side of the prosthesis was lower by a mean of 7% than that on the control side. The difference was greatest in the area of the calcar and laterally and proximally around the stem. BMD at the metaphyseal and diaphyseal areas were the same as on the contralateral side. The bone loss around the proximal aspect of the stem may be related to the proximal porous coating. It is concluded that stress shielding can be diminished by appropriate design of the femoral component.
机译:在24名使用短解剖型股骨干进行全髋关节置换的患者中,使用双能X线吸收法对7年的随访进行了骨矿物质密度(BMD)的测量。将对侧作为对照。假体一侧的BMD平均比对照组降低了7%。差异最大的是在骨的区域以及茎周围的侧面和近端。骨干区和骨干区的骨密度与对侧相同。茎近端周围的骨丢失可能与近端多孔涂层有关。结论是,通过适当设计股骨组件可以减少应力屏蔽。