We demonstrate a blood analysis routine by observing red blood cells through light and digital holographic microscopy in a microfluidic channel. With this setup a determination of red blood cell (RBC) concentration, the mean corpuscular volume (MCV), and corpuscular hemoglobin concentration mean (CHCM) is feasible. Cell count variations in between measurements differed by 2.47% with a deviation of −0.26×106 μL to the reference value obtained from the Siemens ADVIA 2120i. Measured MCV values varied by 2.25% and CHCM values by 3.78% compared to the reference ADVIA measurement. Our results suggest that the combination of optical analysis with microfluidics handling provides a promising new approach to red blood cell counts.
机译:通过在微流体通道中通过光和数字全息显微镜观察红细胞来证明血液分析常规。通过该设置,红细胞(RBC)浓度的测定,平均肉毒体积(MCV)和肉甘油血红蛋白浓度(CHCM)是可行的。测量之间的细胞计数变化不同2.47%,其偏差为-0.26×106μl,从Siemens Advia 2120i获得的参考值。与参考advia测量相比,测量的MCV值通过3.78%而变化2.25%和CHCM值。我们的研究结果表明,使用微流体处理的光学分析的结合为红细胞计数提供了有希望的新方法。