摘要

Oxinium implants are composed of a zirconium alloy with a hard ceramic surface formed by oxidization of the outer layer. This material has been shown to be an effective bearing surface for total joint arthroplasty and an alternative material for use in patients with metal hypersensitivity. Reports exist of metallosis due to unintended wear of Oxinium components from multiple mechanisms including polyethylene liner dissociation and joint instability. This metallosis creates a distinct appearance on radiographic imaging similar to that of an arthrogram. We report 2 cases of metallosis and describe the characteristic radiographic appearance of failed oxidized zirconium implants.
机译:氧植入植入物由锆合金组成,锆合金具有通过外层的氧化形成的硬陶瓷表面。该材料已被证明是总关节置换术的有效轴承表面和用于金属超敏反应患者的替代材料。由于来自多种机制的氧肺组分的意外磨损,存在金属异位的报告,包括聚乙烯衬里解离和关节不稳定性。这种金属异常在类似于αGrogram的射线照相成像上产生明显的外观。我们报告了2例金属异位,描述了失败的氧化锆植入物的特征射线照相外观。