Oxidized Zirconium Total Knee Arthroplasty Implant Failure in a Patient With Knee Instability

机译:氧化锆总膝关节置换术植入物在膝关节不稳定的患者中植入物

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Oxidized zirconium was introduced as an alternative bearing surface to decrease polyethylene wear and reduce aseptic mechanical failure of hip and knee arthroplasties. Oxidized zirconium combines the strength of a metal with wear properties of ceramic, proposing increased survivorship of implant components, and possible decreased rate of revision. Despite a harder surface than cobalt-chromium, the substrate of zirconium is a softer metal. Although uncommon, prior reports have described accelerated wear and severe metallosis after femoral head dislocation in oxidized zirconium total hip arthroplasty; however, this has not been described in total knee arthroplasty. We report a case of an oxidized zirconium total knee arthroplasty failure in a patient with knee instability. This is the first report of catastrophic failure of an oxidized zirconium total knee arthroplasty implant.
机译:将氧化锆作为替代轴承表面引入,以降低聚乙烯磨损,减少髋关节和膝关节育化术的无菌机械失效。氧化锆结合了陶瓷磨损性能的金属强度,提出增加植入物组分的存活,并且可能降低的修订率降低。尽管表面较硬,但含有钴 - 铬的表面,锆的基材是更柔软的金属。虽然罕见,先前的报道已经描述了氧化锆总髋关节成形术中股骨头脱位后的加速磨损和严重金属异位;然而,这尚未在全膝关节形成术中描述。我们在患者患者中举报了含有膝盖不稳定性的患者氧化锆总膝关节置换术失败的情况。这是氧化锆总膝关节形成术植入物灾难性失效的第一报告。