Three‐way comparison of methods for the measurement of the erythrocyte sedimentation rate

机译:红细胞沉降率测量方法的三向比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

An erythrocyte sedimentation rate (ESR) is a nonspecific sickness index, which is not diagnostic of any particular disease, but which when elevated may indicate the presence of inflammation, infection, rheumatologic disease, or neoplasm. Technological advances continue to evolve to make this old test to conform to requirements of a modern analytical laboratory. In this evaluation, two new semi‐automated ESR measuring systems, the HumaSed and ESR‐Auto Plus (EAP), were compared with the Westergren method with regard to the accuracy and precision of ESR measurement, sample stability, and interference of various substances with the ESR assay. Samples from 125 patients were analyzed with the three methods and the results compared using linear regression and Bland–Altman plots. The mean ESR values of the HumaSed (32.10±4.86) and EAP (38.09±5.33) were comparable to that of the Westergren (31.54±4.94). The high correlation coefficients of 0.910–0.96 and the Bland–Altman scatter plots revealed good association and agreement between the three methods. Bias between the three methods was small and the imprecision was within acceptable limits. ESR analysis beyond 4 hr was found to be unacceptable owing to sample instability. There was bilirubin and lipid but not heparin interference in the two automated systems. Overall, two automated analyzers were found to be fast, reliable, standardized, simplified, and safe instruments with accuracy and precision for ESR measurement comparable to the Westergren. J. Clin. Lab. Anal. 22:346–352, 2008. © 2008 Wiley‐Liss, Inc.
机译:红细胞沉降率(ESR)是一种非特异性疾病指数,不能诊断任何特定疾病,但升高时可能表明存在炎症,感染,风湿性疾病或肿瘤。技术进步不断发展,以使这种旧测试符合现代分析实验室的要求。在此评估中,将两个新的半自动ESR测量系统HumaSed和ESR-Auto Plus(EAP)与Westergren方法进行了比较,以了解ESR测量的准确性和精密度,样品稳定性以及各种物质的干扰。 ESR分析。用三种方法分析了来自125位患者的样本,并使用线性回归和Bland–Altman图对结果进行了比较。 HumaSed(32.10±4.86)和EAP(38.09±5.33)的平均ESR值与Westergren(31.54±4.94)相当。 0.910-0.96的高相关系数和布兰德-奥特曼散布图显示出这三种方法之间的良好关联和一致性。三种方法之间的偏差很小,并且不精确度在可接受的范围内。由于样品的不稳定性,超过4小时的ESR分析被认为是不可接受的。在两个自动化系统中存在胆红素和脂质,但没有肝素干扰。总体而言,发现有两台自动化分析仪是快速,可靠,标准化,简化和安全的仪器,其准确度和精密度可与Westergren媲美。 J.临床实验室肛门22:346–352,2008年。©2008 Wiley-Liss,Inc.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号