In the current study, Kovač and colleagues [ ] found an increased risk of THA implant fracture when orthopaedic surgeons use modular necks and concluded that these implants should not be used. I concur with their conclusion, and believe it is fair to ask why this innovation ever was used at all. My group and I used a modular neck in my own implant design in 1982 because we felt it would offer superiority in restoration of hip length and offset. I was warned then by my mentor Dr. Chit Ranawat that the more moving parts added to a design, the greater the likelihood of one of those parts failing. Thanks to that advice, I abandoned modular necks in my subsequent designs. I also learned through experience that good surgery is superior to implant modifications in terms of compensating for anatomic variations. After using computer-assisted navigation for complex THAs for 17 years, I learned that most of the femoral anteversion in dysplastic hips is caused by antetorsion of the head (head-neck angle). And with a neck cut at 10 mm above the lesser trochanter, I have never observed stem anteversion greater than 30°. Because I use cemented stems in some hips, and combined anteversion (lower anteversion in cup) in all, I have never used modularity for dysplastic hips.
机译:在当前的研究中,Kovač及其同事[]发现整形外科医生使用模块化颈部时THA植入物骨折的风险增加,并得出结论,不应使用这些植入物。我同意他们的结论,并相信公平地问为什么要使用这种创新。我和我的小组于1982年在自己的植入物设计中使用了模块化的颈部,因为我们认为它在恢复髋关节长度和偏移方面具有优势。我的导师Chit Ranawat博士当时警告我,设计中添加的运动部件越多,这些部件之一发生故障的可能性就越大。由于这一建议,我在后续设计中放弃了模块化的颈部。我还从经验中学到,就补偿解剖学差异而言,良好的手术要优于植入物的改良。在对复杂的THA使用计算机辅助导航17年后,我了解到,发育不良的髋关节大多数股骨前倾是由头部前倾(头颈部角度)引起的。而且,在小转子上方10毫米处切过脖子时,我从未观察到茎前倾大于30度。因为我在某些臀部使用了水泥杆,并且全部使用了前倾(杯中的前倾度较低),所以我从来没有对发育不良的臀部使用过模块化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号