pH-Mediated field-amplified sample stacking of pharmaceutical cations in high-ionic strength samples

机译:pH值介导的高离子强度样品中药物阳离子的现场放大样品堆叠

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Capillary electrophoretic separation of samples of physiological origin typically have both poor resolution and efficiency due to destacking. We have previously reported a stacking method for concentration of catecholamines in artifical dialysate, or Ringer’s solution. However, pH-mediated sample stacking of other cations has not been investigated. In this report, pH-mediated stacking has been extended to eletripan, dofetilide, doxazosin, sildenafil, UK-103,320, UK-202,581, and CP-122,288. These compounds were chosen without prior structural screening except that they were cationic at the pH of our background electrolyte (BGE). Capillary electrophoretic behavior of samples in BGE is compared with those of samples in Ringer’s solution with and without pH-mediated acid stacking. Results indicate that the peak heights and efficiencies for acid-stacked samples are increased compared to the unstacked samples in Ringer’s solution or BGE. For example, the peak efficiencies for 5 s injections of eletriptan in BGE and Ringer’s solution are 138 000 and 72000 plates, respectively. In contrast, a 10 s injection of eletriptan followed by acid injection for 16 s produces a peak with 246 000 plates. Evaluation of the stacking effect was performed by comparison of the peak height at similar peak efficiencies for samples in Ringer’s solution with and without stacking. Using this method, pH-mediated acid stacking provides a 10- to 27-fold sensitivity enhancement for the seven cations.
机译:生理性样品的毛细管电泳分离通常会由于拆栈而导致分离度和效率低下。我们之前曾报道过一种堆积方法,用于在人造透析液或林格液中浓缩儿茶酚胺。但是,尚未研究pH介导的其他阳离子样品的堆积。在此报告中,pH介导的叠加作用已扩展到依曲坦,多非利特,多沙唑嗪,西地那非,UK-103,320,UK-202,581和CP-122,288。选择这些化合物时无需事先进行结构筛选,只是它们在我们背景电解质(BGE)的pH值上是阳离子。将BGE中样品的毛细管电泳行为与在有或没有pH介导的酸堆积的情况下在林格溶液中的样品进行了比较。结果表明,与在林格溶液或BGE中未堆叠的样品相比,酸堆叠的样品的峰高和效率有所提高。例如,在BGE和Ringer溶液中注射伊曲肽5 s的最高效率分别为138 000和72000板。相反,依曲曲坦注射10 s,然后酸注射16 s产生的峰为246 000个板。堆叠效果的评估是通过比较有和没有堆叠的林格氏溶液中样品在相似峰效率下的峰高进行的。使用这种方法,pH介导的酸堆积对七个阳离子的灵敏度提高了10到27倍。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号