Rapid-Motion-Perception Based Cardiac Navigators: Using the High Flow Blood Volume as a Marker for the Position of the Heart

机译:基于快速运动感知的心脏导航仪:使用高流量血量作为心脏位置的标记

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Navigators have been developed as one of the many approaches to reducing motion artifacts due to respiration. A typical navigator approach applies a pencil-beam style profile crossing the diaphragm to track the superior-inferior (SI) motion of the diaphragm, and subsequently applying correlations to determine the heart location. This approach necessitates a priori knowledge of the correlation coefficients between heart and diaphragm motion, a variable parameter among patients. This paper presents an alternative navigator method based on Rapid Motion Perception (RaMP). This method acquires profiles of the ventricular blood volume based on its high flow velocity. The position of the blood volume is a direct representation of the position of the heart. This method allows cardiac navigation in two orthogonal directions simultaneously, eliminates the need to obtain correlations to the diaphragm motion, and increases tracking reliability for individual patients. A prospective version of RaMP navigators has been implemented on a clinical 1.5 T scanner, and preliminary tests on human volunteers show that this method can successfully track the heart position over the entire respiratory period. This navigation scheme is tested for predicting superior-inferior and anterior-posterior (AP) motion of the heart for breath-hold and free breathing conditions. Bland-Altman plots comparing the motion predicted by the navigators and that computed from single-shot images immediately following the navigators, show that the accuracy of this method is ±1.43 mm in the SI direction and ±0.84 mm in the AP direction. The RaMP navigator is suited for real-time tracking of the bulk translational motion of the heart.
机译:导航器已被开发为减少由于呼吸而引起的运动伪影的众多方法之一。一种典型的导航器方法是应用铅笔束样式的轮廓横穿隔膜,以跟踪隔膜的上下(SI)运动,然后应用相关性来确定心脏位置。这种方法需要先验知识心脏和diaphragm肌运动之间的相关系数,这是患者之间的可变参数。本文提出了一种基于快速运动感知(RaMP)的导航方法。该方法基于其高流速获取心室血容量的轮廓。血容量的位置直接表示心脏的位置。该方法允许在两个正交方向上同时进行心脏导航,消除了获得与the肌运动相关性的需要,并提高了对单个患者的跟踪可靠性。 RaMP导航仪的前瞻性版本已在临床1.5 T扫描仪上实现,对人类志愿者的初步测试表明,该方法可以在整个呼吸周期内成功跟踪心脏位置。测试了此导航方案,以预测屏气和自由呼吸条件下心脏的上下左右运动。 Bland-Altman图将导航器预测的运动与从导航器之后的单次图像计算出的运动进行比较,表明该方法的精度在SI方向为±1.43 mm,在AP方向为±0.84 mm。 RaMP导航仪适用于实时跟踪心脏的整体平移运动。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号