Evidence that the National Committee for Clinical Laboratory Standards disk test is less sensitive than the screen plate for detection of low-expression-class methicillin-resistant Staphylococcus aureus.

机译:有证据表明美国临床实验室标准委员会的磁盘测试对检测低表达级别耐甲氧西林的金黄色葡萄球菌的敏感性不如筛板。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A low-expression-class methicillin-resistant Staphylococcus aureus strain (strain 9302-2) was sent to 207 laboratories as part of the bacteriology component of the Laboratory Proficiency Testing Program of Ontario. An incorrect (susceptible) result was reported by 16 of 76 (21%) of laboratories that used the National Committee for Clinical Laboratory standards disk test, whereas 1 of 104 laboratories that used other methods reported an incorrect result (P < 0.05). Experiments showed discrepancies in the disk test results given by Mueller-Hinton agars from three manufacturers. We advise that laboratories should use a low expression-class methicillin-resistant S. aureus isolate as a control for the National Committee for Clinical Laboratory Standards disk test.
机译:作为安大略省实验室能力测试计划细菌学组成部分的一部分,将低表达水平的耐甲氧西林金黄色葡萄球菌菌株(菌株9302-2)发送到207个实验室。使用国家临床实验室委员会标准磁盘测试的76个实验室中有16个(21%)报告了不正确的(敏感)结果,而使用其他方法的104个实验室中有1个报告了不正确的结果(P <0.05)。实验表明,来自三个制造商的Mueller-Hinton琼脂的圆盘测试结果存在差异。我们建议实验室应使用低表达级别耐甲氧西林的金黄色葡萄球菌作为国家临床实验室标准委员会磁盘测试的对照。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号