首页> 美国卫生研究院文献>other >Estimates of Body Fat in Children by Hologic QDR-2000 and QDR-4500A Dual-Energy X-ray Absorptiometers Compared With Deuterium Dilution
【2h】

Estimates of Body Fat in Children by Hologic QDR-2000 and QDR-4500A Dual-Energy X-ray Absorptiometers Compared With Deuterium Dilution

机译:用Hologic QDR-2000和QDR-4500A双能X射线吸收仪与氘稀释法比较儿童体内脂肪

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We evaluated the accuracy with which the Hologic QDR-4500A and QDR-2000 densitometers measure fat mass (FM) in 95 children. FM was derived from total body water measured by deuterium dilution (DD) in all children, by QDR-4500A in 50, and by QDR-2000 in 45 children. Compared with DD, both instruments underestimated FM (QDR-4500A: 3.35 ± 2.5 kg, P < 0.0001; QDR-2000: 1.05 ± 1.5 kg, P < 0.0001). Both absorptiometers showed magnitude biases relative to DD (QDR-4500: r = +0.70, P < 0.001; QDR-2000: r = −0.51, P < 0.001). We conclude that neither densitometer is equivalent to DD for estimation of children’s FM. The QDR-4500A’s current calibration seems to provide an even greater underestimate of FM than the QDR-2000.
机译:我们评估了Hologic QDR-4500A和QDR-2000密度计在95名儿童中测量脂肪量(FM)的准确性。 FM来自所有儿童的总体内水分,这些水分是通过氘稀释(DD),50名儿童的QDR-4500A和45名儿童的QDR-2000测得的。与DD相比,两种仪器都低估了FM(QDR-4500A:3.35±2.5千克,P <0.0001; QDR-2000:1.05±1.5千克,P <0.0001)。两种吸光度计均显示相对于DD的幅度偏差(QDR-4500:r = +0.70,P <0.001; QDR-2000:r = -0.51,P <0.001)。我们得出的结论是,任何一种密度计都不能等同于DD来估计儿童的FM。与QDR-2000相比,QDR-4500A的当前校准似乎提供了更大的FM低估。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号