Estimates of body fat in children by Hologic QDR-2000 and QDR-4500A dual-energy X-ray absorptiometers compared with deuterium dilution.

机译:通过Hologic QDR-2000和QDR-4500A双能X射线吸收仪与氘稀释相比,可估算儿童体内的脂肪。

获取原文
获取原文并翻译 | 示例

摘要

We evaluated the accuracy with which the Hologic QDR-4500A and QDR-2000 densitometers measure fat mass (FM) in 95 children. FM was derived from total body water measured by deuterium dilution (DD) in all children, by QDR-4500A in 50, and by QDR-2000 in 45 children. Compared with DD, both instruments underestimated FM (QDR-4500A: 3.35 +/- 2.5 kg, P < 0.0001; QDR-2000: 1.05 +/- 1.5 kg, P < 0.0001). Both absorptiometers showed magnitude biases relative to DD (QDR-4500: r = +0.70, P < 0.001; QDR-2000: r = -0.51, P < 0.001). We conclude that neither densitometer is equivalent to DD for estimation of children's FM. The QDR-4500A's current calibration seems to provide an even greater underestimate of FM than the QDR-2000.
机译:我们评估了Hologic QDR-4500A和QDR-2000密度计在95名儿童中测量脂肪量(FM)的准确性。 FM来自所有儿童的总体内水分,这些水分是通过氘稀释(DD),50名儿童的QDR-4500A和45名儿童的QDR-2000测得的。与DD相比,两台仪器都低估了FM(QDR-4500A:3.35 +/- 2.5千克,P <0.0001; QDR-2000:1.05 +/- 1.5千克,P <0.0001)。两种吸光度计均显示相对于DD的幅度偏差(QDR-4500:r = +0.70,P <0.001; QDR-2000:r = -0.51,P <0.001)。我们得出的结论是,两种光密度计都不能等同于DD来估计儿童的FM。与QDR-2000相比,QDR-4500A的当前校准似乎提供了更大的FM低估。