首页> 美国卫生研究院文献>other >A compact 7-cell Si-drift detector module for high-count rate X-ray spectroscopy
【2h】

A compact 7-cell Si-drift detector module for high-count rate X-ray spectroscopy

机译:紧凑的7单元硅漂移检测器模块用于高计数率X射线光谱学

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A new Si-drift detector module for fast X-ray spectroscopy experiments was developed and realized. The Peltier-cooled module comprises a sensor with 7 × 7-mm2 active area, an integrated circuit for amplification, shaping and detection, storage, and derandomized readout of signal pulses in parallel, and amplifiers for line driving. The compactness and hexagonal shape of the module with a wrench size of 16mm allow very short distances to the specimen and multi-module arrangements. The power dissipation is 186mW. At a shaper peaking time of 190 ns and an integration time of 450 ns an electronic rms noise of ~11 electrons was achieved. When operated at 7 °C, FWHM line widths around 260 and 460 eV (Cu-Kα) were obtained at low rates and at sum-count rates of 1.7 MHz, respectively. The peak shift is below 1% for a broad range of count rates. At 1.7-MHz sum-count rate the throughput loss amounts to 30%.
机译:开发并实现了用于快速X射线光谱实验的新型Si漂移检测器模块。珀尔帖冷却模块包括一个具有7×7mm 2 有源区域的传感器,一个用于并行放大,整形和检测,存储和非随机读出信号脉冲的集成电路,以及用于线路的放大器驾驶。扳手尺寸为16mm的模块的紧凑性和六边形形状使得到样品和多模块装置的距离非常短。功耗为186mW。在整形器的峰值时间为190 ns,积分时间为450 ns时,获得的电子均方根噪声约为11个电子。当在7°C下工作时,分别以低速率和1.7 MHz的总计数速率获得了大约260和460 eV(Cu-Kα)的FWHM线宽。对于广泛的计数率,峰值漂移低于1%。在1.7MHz的总计数速率下,吞吐量损失总计为30%。