Functional Evaluation of the Reusable JuniorSTAR® Half-Unit Insulin Pen

机译:可重复使用的JuniorSTAR®半单位胰岛素笔的功能评估

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Background: The functional performance of the JuniorSTAR® (Sanofi, Paris, France) half-unit insulin pen was evaluated through a series of specific objective tests to assess the dose accuracy, pen weight, injection force, and dialing torque. >Method: Pens (n = 60) were tested under standard atmospheric conditions with 3 different types of insulins manufactured by Sanofi (insulin glargine, insulin glulisine, and biphasic insulin isophane). The dose accuracy was tested according to the ISO 11608-1:2012 standards. Injection doses of 0.010, 0.155, and 0.300 ml were evaluated. For mean weight evaluation, the pens without the cartridge were weighed on precision balances. The injection force was measured using a texture analyzer and the dialing torque was measured using a torque meter. >Results: JuniorSTAR met the ISO 11608-1:2012 criteria for dose accuracy as all the delivered doses were within the predefined limits for all types of insulin tested. The mean weight of the JuniorSTAR pen was 33.4 g (SD = 0.075). The mean injection force was 6.0 N (SD = 0.8), 4.3 N (SD = 0.4), and 5.1 N (SD = 0.6) for insulin glargine, insulin glulisine, and biphasic insulin isophane, respectively. The mean dialing torque was 5.09 Ncm (SD = 0.29) and 5.88 Ncm (SD = 0.53) for setting and correcting a dose, respectively. >Conclusions: Together with results from a previously reported usability survey, these results show that the JuniorSTAR reusable, half-unit pen is a lightweight and accurate device for insulin delivery with a dialing torque and injection force suitable for young people with type 1 diabetes.
机译:>背景:通过一系列具体的客观测试评估了JuniorSTAR ®(法国巴黎的萨诺菲)半单位胰岛素笔的功能性能,以评估剂量的准确性,笔的重量,注射力和拨号扭矩。 >方法:在标准大气条件下,使用赛诺菲生产的3种不同类型的胰岛素(甘精胰岛素,甘精胰岛素和异相双相胰岛素)对笔(n = 60)进行了测试。剂量精度根据ISO 11608-1:2012标准进行测试。评估了0.010、0.155和0.300 ml的注射剂量。为了进行平均重量评估,将没有墨盒的笔在精密天平上称重。使用质地分析仪测量注射力,并使用扭矩计测量拨度扭矩。 >结果:JuniorSTAR的剂量准确性符合ISO 11608-1:2012标准,因为所有递送的剂量均在所有测试的胰岛素类型的预定限制内。 JuniorSTAR笔的平均重量为33.4克(SD = 0.075)。甘精胰岛素,甘精胰岛素和双相异硫烷的平均注射力分别为6.0 N(SD = 0.8),4.3 N(SD = 0.4)和5.1 N(SD = 0.6)。设定和校正剂量的平均拨号扭矩分别为5.09 Ncm(SD = 0.29)和5.88 Ncm(SD = 0.53)。 >结论:这些结果与先前报告的可用性调查的结果一起表明,JuniorSTAR可重复使用的半单元笔是一种轻巧而精确的胰岛素输送装置,其拨号扭矩和注射力适合于年轻人1型糖尿病患者。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号