Tissue elasticity is related to pathology and therefore has important medical applications. Radiation force from a focused ultrasound beam has been used to produce shear waves in tissues for shear wave speed and tissue elasticity measurements. The feasibility of shear wave speed measurement using radiation force for an unfocused ultrasound beam is demonstrated in this study with a linear and a curved array transducer. Consistent measurement of shear wave speed was achieved over a relatively long axial extent (z = 10-40 mm for the linear array, and z = 15-60 mm for the curved array) in 3 calibrated phantoms with different shear moduli. In vivo measurements on the biceps of a healthy volunteer show consistent increase of shear wave speed for the biceps under 0, 1, 2, and 3 kg loading. Advantages and limitations of unfocused push are discussed.
机译:组织弹性与病理学有关,因此具有重要的医疗应用。来自聚焦超声波梁的辐射力已被用于在用于剪切波速和组织弹性测量的组织中产生剪切波。在该研究中,用线性和弯曲阵列换能器对使用未聚焦超声波束的辐射力进行剪力波速度测量的可行性。在具有不同剪切模量的3个校准的模型中,在相对长的轴向范围内(用于线性阵列的Z = 10-40mm,Z = 15-60mm的Z = 15-60mm),实现了一致的剪切波速度的测量。在健康志愿者的二头肌的体内测量中,展示了0,1,2和3kg负载下的二头肌的剪力速度一致增加。讨论了未聚焦推动的优点和局限性。