摘要

Corneal grafts for Descemet’s Stripping Automated Endothelial Keratoplasty are commonly prepared using mechanical microkeratomes. However, the cuts produced in such way render corneal lenticules that are thinner centrally than peripherally, thus inducing a hyperopic shift. Here we describe a novel device for preparing donor corneal grafts, in which a single low-energy femtosecond laser system is used as both a light source for optical coherence tomography and for cutting the graft illuminating from the endothelial side. The same laser is first utilized to obtain three-dimensional optical coherence tomography images of the donor tissue for guiding the dissection and obtaining grafts of uniform thickness with no applanation or contact. This device allows an optimal procedure for preparing consistently thin posterior grafts for transplantation.
机译:通常使用机械微角膜刀制备用于Descemet剥离自动内皮角膜移植术的角膜移植物。然而,以这种方式产生的切口使得角膜微透镜中央比周围更细,从而引起远视移位。在这里,我们描述了一种用于制备供体角膜移植物的新型装置,其中单个低能量飞秒激光系统既用作光学相干层析成像的光源,又用于从内皮侧切割移植物照明。首先使用相同的激光获得供体组织的三维光学相干断层扫描图像,以引导解剖并获得厚度均匀,没有压平或接触的移植物。该设备可为准备一致的后壁移植物进行移植提供最佳程序。