Dose and detectability improvements with high energy phase sensitive x-ray imaging in comparison to low energy conventional imaging

机译:与低能常规成像相比高能相敏X射线成像的剂量和可检测性得到了改善

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The objective of this study was to demonstrate the potential benefits of using high energy x-rays for phase sensitive breast imaging through a comparison with conventional mammography imaging. We compared images of a contrast-detail (CD) phantom acquired on a prototype phase sensitive x-ray imaging system with images acquired on a commercial flat panel digital mammography unit. The phase contrast images were acquired using a micro-focus x-ray source with a 50 μm focal spot at 120 kVp and 4.5 mAs, with a magnification factor of 2.46 and a 50 μm pixel pitch. A phase attenuation duality (PAD)-based phase retrieval algorithm that requires only a single phase contrast image was applied. Conventional digital mammography images were acquired at 27 kVp, 131 mAs and 28 kVp, 54 mAs. For the same radiation dose, both the observer study and SNR/FOM comparisons indicated a large improvement by the phase retrieved image as compared to the clinical system for the larger disk sizes, but the improvement was not enough to detect the smallest disks. Compared to the double dose image acquired with the clinical system, the observer study also indicated that the phase retrieved image provided improved detection capabilities for all disk sizes except the smallest disks. Thus the SNR improvement provided by phase contrast imaging is not yet enough to offset the noise reduction provided by the clinical system at the doubled dose level. However, the potential demonstrated by this study for high energy phase sensitive x-ray imaging to improve lesion detection and reduce radiation dose in mammography warrants further investigation of this technique.
机译:这项研究的目的是通过与传统的乳房X线摄影术成像进行比较,证明使用高能X射线进行相敏乳房成像的潜在益处。我们比较了在原型相敏X射线成像系统上获得的对比度细节(CD)体模的图像与在商业平板数字乳腺摄影设备上获得的图像。使用具有50μm焦点,120 kVp和4.5 mAs的微焦点X射线源获取相衬图像,放大倍率为2.46,像素间距为50μm。基于相位衰减对偶性(PAD)的相位检索算法仅需要单个相位对比图像。在27 kVp,131 mAs和28 kVp,54 mAs下采集常规数字乳房X线照片。对于相同的辐射剂量,观察者研究和SNR / FOM比较均表明,相较于较大盘片的临床系统,相位检索图像有很大改进,但这种改进不足以检测最小的盘片。与临床系统获取的双剂量图像相比,观察者研究还表明,相图检索的图像提供了对除最小磁盘以外的所有磁盘大小的改进的检测能力。因此,相衬成像所提供的SNR改善还不足以抵消临床系统在两倍剂量水平下所产生的降噪效果。但是,这项研究显示出的高能相敏X射线成像技术有望改善乳腺摄影中的病灶检测并减少放射剂量,因此有必要对该技术进行进一步研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号