Accuracy of Image Guidance Using Free-Breathing Cone-Beam Computed Tomography for Stereotactic Lung Radiotherapy

机译:自由呼吸锥形束计算机断层扫描在立体定向肺放疗中的图像制导精度

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Movement of the target object during cone-beam computed tomography (CBCT) leads to motion blurring artifacts. The accuracy of manual image matching in image-guided radiotherapy depends on the image quality. We aimed to assess the accuracy of target position localization using free-breathing CBCT during stereotactic lung radiotherapy. The Vero4DRT linear accelerator device was used for the examinations. Reference point discrepancies between the MV X-ray beam and the CBCT system were calculated using a phantom device with a centrally mounted steel ball. The precision of manual image matching between the CBCT and the averaged intensity (AI) images restructured from four-dimensional CT (4DCT) was estimated with a respiratory motion phantom, as determined in evaluations by five independent operators. Reference point discrepancies between the MV X-ray beam and the CBCT image-guidance systems, categorized as left-right (LR), anterior-posterior (AP), and superior-inferior (SI), were 0.33 ± 0.09, 0.16 ± 0.07, and 0.05 ± 0.04 mm, respectively. The LR, AP, and SI values for residual errors from manual image matching were -0.03 ± 0.22, 0.07 ± 0.25, and -0.79 ± 0.68 mm, respectively. The accuracy of target position localization using the Vero4DRT system in our center was 1.07 ± 1.23 mm (2 SD). This study experimentally demonstrated the sufficient level of geometric accuracy using the free-breathing CBCT and the image-guidance system mounted on the Vero4DRT. However, the inter-observer variation and systematic localization error of image matching substantially affected the overall geometric accuracy. Therefore, when using the free-breathing CBCT images, careful consideration of image matching is especially important.
机译:锥束计算机断层扫描(CBCT)期间目标对象的移动会导致运动模糊伪像。图像引导放射治疗中手动图像匹配的准确性取决于图像质量。我们旨在评估立体定向肺放疗期间使用自由呼吸CBCT进行目标位置定位的准确性。检查使用了Vero4DRT线性加速器设备。 MV X射线束和CBCT系统之间的参考点差异是使用带有中央安装钢球的幻像设备计算的。由五个独立操作员的评估确定,使用呼吸运动体模估计了CBCT与从四维CT(4DCT)重构的平均强度(AI)图像之间的手动图像匹配精度。 MV X射线束与CBCT图像引导系统之间的参考点差异为0.33±0.09、0.16±0.07,分别为左右(LR),前后(AP)和上下(SI)和0.05±0.04 mm。手动图像匹配带来的残留误差的LR,AP和SI值分别为-0.03±0.22、0.07±0.25和-0.79±0.68 mm。在我们中心使用Vero4DRT系统进行目标位置定位的精度为1.07±1.23 mm(2 SD)。这项研究通过使用自由呼吸的CBCT和安装在Vero4DRT上的图像引导系统,通过实验证明了足够的几何精度。但是,观察者之间的差异和图像匹配的系统定位误差严重影响了总体几何精度。因此,在使用自由呼吸的CBCT图像时,仔细考虑图像匹配尤为重要。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号