首页> 美国卫生研究院文献>other >An Interface for the Direct Coupling of Small Liquid Samples to AMS
【2h】

An Interface for the Direct Coupling of Small Liquid Samples to AMS

机译:直接将小量液体样品耦合到AMS的接口

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We describe the moving wire interface attached to the 1-MV AMS system at LLNL’s Center for Accelerator Mass Spectrometry for the analysis of nonvolatile liquid samples as either discrete drops or from the direct output of biochemical separatory instrumentation, such as high-performance liquid chromatography. Discrete samples containing at least a few 10s of nanograms of carbon and as little as 50 zmol 14C can be measured with a 3–5% precision in a few minutes. The dynamic range of our system spans approximately 3 orders in magnitude. Sample to sample memory is minimized by the use of fresh targets for each discrete sample or by minimizing the amount of carbon present in a peak generated by an HPLC containing a significant amount of 14C. Liquid Sample AMS provides a new technology to expand our biomedical AMS program by enabling the capability to measure low-level biochemicals in extremely small samples that would otherwise be inaccessible.
机译:我们在LLNL的加速器质谱中心描述了连接到1-MV AMS系统的可动线接口,用于分析非挥发性液体样品的离散滴或直接从生化分离仪器(如高效液相色谱法)中输出。可以在几分钟内以3–5%的精度测量至少包含几十个纳克碳和低至50 zmol 14 C的离散样品。我们系统的动态范围跨度约为3个数量级。通过为每个离散样品使用新的靶标或通过使包含大量 14 C的HPLC产生的峰中存在的碳量最小化,可以最大程度地减少样品间的记忆。液体样品AMS提供了一项新技术,可扩展我们的生物医学AMS计划,使其具有测量极少量样品中低水平生化试剂的能力,否则这些样品将无法获得。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号