The use of extended dose range film for dosimetric calibration of a scanning liquid‐filled ionization chamber electronic portal imaging device

机译:使用扩大剂量范围的膜对扫描液体填充电离室电子门成像设备进行剂量学校准

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A scanning liquid‐filled ionization chamber electronic portal imaging device (SLIC‐EPID) and extended dose range (EDR2) films were used to evaluate transmitted dose profiles for homogeneous and inhomogeneous phantoms. Calibrated ionization chamber measurements were used to convert the pixel values acquired from the electronic portal images to dose. Because SLIC‐EPID was developed to have a uniform response for all liquid ionization chambers, the off‐axis dose values were reconstructed using a correction factor matrix, defined as the ratio of the relative EDR2 film and the corresponding EPID dose values measured in air. The transmitted dose distributions in the EPID detector layer were also modeled using a Pinnacle3 treatment planning system (TPS: Philips Radiation Oncology Systems, Milpitas, CA). The gamma function algorithm was then used to assess agreement between transmitted dose distributions measured using a SLIC‐EPID and EDR2 film, and those calculated using the TPS. For homogenous and inhomogeneous phantoms, more than 90% agreement was achieved using gamma criteria of 2% and 3 mm and 3% and 2.5 mm respectively. Our results indicate that the calibration procedure proposed in the present study should be performed if SLIC‐EPID is to be used as a reliable two‐dimensional transmitted dosimeter for clinical purposes.PACS numbers: 87.53.Tf, 87.53.Oq
机译:使用扫描液体填充电离室电子门成像设备(SLIC-EPID)和扩展剂量范围(EDR2)胶片来评估均匀和不均匀体模的透射剂量分布。校准的电离室测量值用于将从电子门禁图像获取的像素值转换为剂量。由于SLIC-EPID被开发为对所有液体电离室具有一致的响应,因此使用校正因子矩阵重建离轴剂量值,该校正因子矩阵定义为相对EDR2膜与空气中测得的相应EPID剂量值之比。还使用Pinnacle 3 治疗计划系统(TPS:Philips Radiation Oncology Systems,Milpitas,CA)对EPID检测器层中的透射剂量分布进行了建模。然后使用伽马函数算法评估使用SLIC-EPID和EDR2胶片测量的透射剂量分布与使用TPS计算的透射剂量分布之间的一致性。对于均质和不均质体模,分别使用2%和3 mm,3%和2.5 mm的伽玛标准可以达到90%以上的一致性。我们的结果表明,如果将SLIC-EPID用作临床上可靠的二维透射剂量计,则应执行本研究中提出的校准程序.PACS编号:87.53.Tf,87.53.Oq

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号