Physics testing necessary for program accreditation is rigorously defined by the ACR. This testing is easily applied to most conventional SPECT systems based on gamma camera technology. The inSPira HD is a dedicated head SPECT system based on a rotating dual clamshell design that acquires data in a dual‐spiral geometry. The unique geometry and configuration force alterations of the standard ACR physics testing protocol. Various tests, such as intrinsic planar uniformity and/or resolution, do not apply. The Data Spectrum Deluxe Phantom used for conventional SPECT testing cannot fit in the inSPira HD scanner bore, making (currently) unapproved use of the Small Deluxe SPECT Phantom necessary. Matrix size, collimator type, scanning time, reconstruction method, and attenuation correction were all varied from the typically prescribed ACR instructions. Visible spheres, sphere contrast, visible rod groups, uniformity, and root mean square (RMS) noise were measured. The acquired SPECT images surpassed the minimum ACR requirements for both spatial resolution (9.5 mm spheres resolved) and contrast (6.4 mm rod groups resolved). Sphere contrast was generally high. Integral uniformity was 4% and RMS noise was 1.7%. Noise appeared more correlated than in images from a conventional SPECT scanner. Attenuation‐corrected images produced from direct CT scanning of the phantom and a manufacturer supplied model of the phantom demonstrated negligible differences.PACS numbers: 87.57.C‐, 87.57.uh, 87.63.lj
机译:ACR严格定义了程序认证所需的物理测试。该测试可轻松应用于基于伽马相机技术的大多数常规SPECT系统。 inSPira HD是基于旋转双翻盖设计的专用头SPECT系统,该系统以双螺旋几何形状采集数据。标准ACR物理测试协议的独特几何形状和配置强制更改。各种测试(例如固有的平面均匀性和/或分辨率)均不适用。用于常规SPECT测试的Data Spectrum Deluxe Phantom无法放入inSPira HD扫描仪孔中,因此(当前)未经批准使用Small Deluxe SPECT Phantom是必要的。矩阵尺寸,准直仪类型,扫描时间,重建方法和衰减校正均不同于通常规定的ACR指令。测量了可见球,球对比度,可见棒组,均匀性和均方根(RMS)噪声。对于空间分辨率(分辨出9.5毫米球体)和对比度(分辨出6.4毫米杆组),所获得的SPECT图像均超过了最低ACR要求。球体对比度通常很高。整体均匀度为4%,RMS噪声为1.7%。与传统SPECT扫描仪的图像相比,噪声似乎更相关。由体模直接CT扫描和制造商提供的体模模型产生的经过衰减校正的图像显示的差异可以忽略不计.PACS编号:87.57.C-,87.57.uh,87.63.lj

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号