The capacity of 4DCT to quantify organ motion is beyond conventional 3DCT capability. Local control could be improved. However we are unaware of any reports of organ dose measurements for helical 4DCT imaging. We therefore quantified the radiation doses for helical 4DCT imaging. Organ and tissue dose was measured for thoracic and abdominal 4DCT in helical mode using an adult anthropomorphic phantom. Radiation doses were measured with thermoluminescence dosimeter chips inserted at various anatomical sites on the phantom. For the helical thoracic 4DCT, organ doses were 57.2 mGy for the lung, 76.7 mGy for the thyroids, 48.1 mGy for the breasts, and 10.86 mGy for the colon. The effective doses for male and female phantoms were very similar, with a mean value of 33.1 mSv. For abdominal 4DCT imaging, organ doses were 14.4 mGy for the lung, 0.78 mGy for the thyroids, 9.83 mGy for breasts, and 58.2 mGy for the colon (all obtained by using ICRP 103). We quantified the radiation exposure for thoracic and abdominal helical 4DCT. The doses for helical 4DCT were approximately 1.5 times higher than those for cine 4DCT, however the stepwise image artifact was reduced. 4DCT imaging should be performed with care in order to minimize radiation exposure, but the advantages of 4DCT imaging mandates its incorporation into routine treatment protocols.
机译:4DCT量化器官运动的能力超出了常规3DCT的能力。本地控制可以改善。但是,我们没有任何关于螺旋4DCT成像的器官剂量测量的报告。因此,我们量化了螺旋4DCT成像的辐射剂量。使用成人拟人体模以螺旋模式测量胸和腹部4DCT的器官和组织剂量。用插入在体模上各个解剖部位的热致发光剂量计芯片测量辐射剂量。对于螺旋胸4DCT,器官剂量对肺是57.2 mGy,对甲状腺是76.7 mGy,对乳房是48.1 mGy,对结肠是10.86 mGy。男性和女性幻影的有效剂量非常相似,平均值为33.1 mSv。对于腹部4DCT成像,肺脏器官剂量为14.4 mGy,甲状腺器官剂量为0.78 mGy,乳房器官剂量为9.83 mGy,结肠器官剂量为58.2 mGy(所有剂量均使用ICRP 103获得)。我们量化了胸部和腹部螺旋4DCT的辐射暴露。螺旋4DCT的剂量约为电影4DCT的剂量的1.5倍,但是逐步减少了图像伪影。 4DCT成像应谨慎进行,以最大程度地减少放射线照射,但是4DCT成像的优点要求将其纳入常规治疗方案中。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号