摘要

Magnetic resonance imaging >( MRI) is being rapidly integrated for cancer treatments—such systems are referred to as MRI‐guided radiation therapy (MRIgRT). As the magnet of an MRI scanner is always on, the presence of a strong static magnetic field from the MRI scanner during radiotherapy delivery presents new challenges. One of the challenges is that a personal radiation dosimeter used to estimate the radiation dose deposited in an individual wearing the device must be MR‐safe. No such devices, however, are currently available. In this work we first modified an existing personal dosimeter (by removing a metal clip) to make it MR‐safe and then investigated potential effects of magnetic field on dosimeter readings, i.e., optically stimulated luminescent dosimeter (OSLD) readings. We found that the effect of magnetic field on OSLD sensitivity was within radiation protection tolerance levels. OSLD personal dosimeters can be directly used in conjunction with MRIgRT radiation protection purposes.
机译:磁共振成像>( MRI)正在迅速集成到癌症治疗中,这种系统被称为MRI引导放射治疗(MRIgRT)。由于MRI扫描仪的磁体始终处于打开状态,因此在放射治疗期间MRI扫描仪存在强大的静态磁场会带来新的挑战。挑战之一是用于估计在佩戴设备的个人中沉积的辐射剂量的个人辐射剂量计必须是MR安全的。但是,目前尚无此类设备。在这项工作中,我们首先修改了一个现有的个人剂量计(通过移除金属夹)使其具有MR安全性,然后研究了磁场对剂量计读数的潜在影响,即光学激发发光剂量计(OSLD)读数。我们发现磁场对OSLD灵敏度的影响在辐射防护公差水平之内。 OSLD个人剂量计可直接与MRIgRT辐射防护结合使用。