首页> 外文期刊>Radiation Protection and Environment >Response of CaSO_4: Dy based thermo luminescence dosimeter badge to ~(137)Cs and ~(60)Co radiations and its implications on estimation of personal dose equivalent
【24h】

Response of CaSO_4: Dy based thermo luminescence dosimeter badge to ~(137)Cs and ~(60)Co radiations and its implications on estimation of personal dose equivalent

机译:CaSO_4:Dy基热发光剂量计徽章对〜(137)Cs和〜(60)Co辐射的响应及其对估算个人剂量当量的影响

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

In view of an underestimation of personal dose equivalent due to ~(60)Co radiation when system is calibrated to ~(137)Cs, the effect of the atomic number of the filters used in the badge on the photon beams of 660 keV and 1250 keV was investigated. The low Z-high Z interface created between filters and discs was found to lead to discontinuity in dose at the surface of the disc. This effect was consistently observable due to the 0.8 mm thickness of the disc. It was concluded that under-estimation of dose due to ~(60)Co radiation could be avoided by either reducing the thickness of the dosimeter disc or by a slight modification in the design of the filters.
机译:鉴于将系统校准至〜(137)Cs时由于〜(60)Co辐射导致的个人剂量当量低估了,徽章中使用的滤光片的原子序数对660 keV和1250的光子束的影响对keV进行了研究。发现在过滤器和圆盘之间形成的低Z高Z界面导致圆盘表面剂量不连续。由于光盘的厚度为0.8 mm,因此可以始终观察到这种效果。得出的结论是,可以通过减小剂量计盘的厚度或对过滤器的设计稍加修改来避免由于〜(60)Co辐射引起的剂量低估。

著录项

  • 来源
    《Radiation Protection and Environment》 |2016年第2期|91-95|共5页
  • 作者单位

    TLD-Personnef Monitoring and Services Section, Radiological Physics and Advisory Division, Bhabha Atomic Research Centre, CT and CRS, Anushaktinagar, Mumbai - 400 094, Maharashtra, India;

    TLD-Personnel Monitoring and Services Section, Radiological Physics and Advisory Division, Bhabha Atomic Research Centre, CT and CRS, Anushaktinagar, Mumbai, Maharashtra, India;

    TLD-Personnel Monitoring and Services Section, Radiological Physics and Advisory Division, Bhabha Atomic Research Centre, CT and CRS, Anushaktinagar, Mumbai, Maharashtra, India;

    TLD-Personnel Monitoring and Services Section, Radiological Physics and Advisory Division, Bhabha Atomic Research Centre, CT and CRS, Anushaktinagar, Mumbai, Maharashtra, India;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Calibration; high Z phosphor; individual monitoring;