A 60‐year‐old male experienced an inappropriate shock from an implantable cardioverter‐defibrillator (ICD) because of oversensing of pectoral myopotentials. Battery depletion was also observed, and a generator change was performed. A single‐chamber ICD (VENTAK PRIZM II 1860) was changed to a new ICD (INCEPTA VR F161). The myopotentials were clearly eliminated by the difference in the band pass filter (PRIZM; 21‐171 Hz, INCEPTA; 20‐85 Hz), but unfortunately, new noise was documented 4 years later. The utility of the bandpass filter for preventing oversensing of myopotentials was observed, but the limitation of its use for long‐term follow‐up was also indicated.
机译:一名60岁男性因胸肌电位过度感应而遭受了植入式心脏复律除颤器(ICD)的不当电击。还观察到电池耗尽,并进行了发电机更换。单腔ICD(VENTAK PRIZM II 1860)更改为新的ICD(INCEPTA VR F161)。通过带通滤波器(PRIZM; 21-171 Hz,INCEPTA; 20-85 Hz)的差异明显消除了肌电势,但不幸的是,四年后才记录下新的噪声。观察到了带通滤波器在防止肌电势过度感应方面的实用性,但也指出了其长期跟踪使用的局限性。