首页> 美国卫生研究院文献>Journal of Applied Crystallography >Investigation of high-pressure planetary ices by cryo-recovery. I. An apparatus for X-ray powder diffraction from 40 to 315 K allowing ‘cold loading’ of samples
【2h】

Investigation of high-pressure planetary ices by cryo-recovery. I. An apparatus for X-ray powder diffraction from 40 to 315 K allowing ‘cold loading’ of samples

机译:通过低温回收研究高压行星冰。 I.用于40到315 K的X射线粉末衍射的设备允许冷装样品

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A low-temperature stage for X-ray powder diffraction in Bragg–Brentano reflection geometry is described. The temperature range covered is 40–315 K, with a temperature stability at the sample within ±0.1 K of the set point. The stage operates by means of a Gifford–McMahon (GM) closed-cycle He refrigerator; it requires no refrigerants and so can run for an extended period (in practice at least 5 d) without intervention by the user. The sample is cooled both by thermal conduction through the metal sample holder and by the presence of He exchange gas, at ambient pressure, within the sample chamber; the consumption of He gas is extremely low, being only 0.1 l min−1 during normal operation. A unique feature of this cold stage is that samples may be introduced into (and removed from) the stage at any temperature in the range 80–300 K, and thus materials which are not stable at room temperature, such as high-pressure phases that are recoverable to ambient pressure after quenching to liquid nitro­gen temperatures, can be readily examined. A further advantage of this arrangement is that, by enabling the use of pre-cooled samples, it greatly reduces the turnaround time when making measurements on a series of specimens at low temperature.
机译:描述了布拉格-布伦塔诺反射几何结构中X射线粉末衍射的低温阶段。涵盖的温度范围为40–315 K,样品的温度稳定性在设定点的±0.1 K之内。该平台通过Gifford-McMahon(GM)封闭式He冰箱运行;它不需要制冷剂,因此可以长时间运行(实际上至少5 dd),而无需用户干预。样品通过金属样品架的热传导和环境室内压力氦气交换气体的存在而冷却。氦气的消耗极低,在正常运行期间仅为0.1 l min -1 。这种低温工作台的独特之处在于,可以在80-300 K范围内的任何温度下将样品引入到工作台中(或从中取出),因此,在室温下不稳定的材料(例如高压相)淬火至液氮温度后,可将其回收至环境压力,可轻松进行检查。这种布置的另一个优点是,通过允许使用预冷的样品,在低温下对一系列样品进行测量时,大大减少了周转时间。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号