>Background/aim: The technique of photoretinoscopy is unique in being able to measure the dynamics of the oculomotor system (ocular accommodation, vergence, and pupil size) remotely (working distance typically 1 metre) and objectively in both eyes simultaneously. The aim of this study was to evaluate clinically the measurement of refractive error by a recent commercial photoretinoscopic device, the PowerRefractor (PlusOptiX, Germany).>Method: The validity and repeatability of the PowerRefractor was compared to: subjective (non-cycloplegic) refraction on 100 adult subjects (mean age 23.8 (SD 5.7) years) and objective autorefraction (Shin-Nippon SRW-5000, Japan) on 150 subjects (20.1 (4.2) years). Repeatability was assessed by examining the differences between autorefractor readings taken from each eye and by re-measuring the objective prescription of 100 eyes at a subsequent session.>Results: On average the PowerRefractor prescription was not significantly different from the subjective refraction, although quite variable (difference +0.05 (0.63) D, p = 0.41) and more negative than the SRW-5000 prescription (by −0.20 (0.72) D, p<0.001). There was no significant bias in the accuracy of the instrument with regard to the type or magnitude of refractive error. The PowerRefractor was found to be repeatable over the prescription range of −8.75D to +4.00D (mean spherical equivalent) examined.>Conclusion: The PowerRefractor is a useful objective screening instrument and because of its remote and rapid measurement of both eyes simultaneously is able to assess the oculomotor response in a variety of unrestricted viewing conditions and patient types.
机译:>背景/目的:光学视网膜检影技术的独特之处在于,它能够远程(通常为1米的工作距离)客观地测量动眼系统的动力学(眼部适应性,散度和瞳孔大小)。眼睛同时。这项研究的目的是通过临床上最新的商用光学视网膜镜设备PowerRefractor(PlusOptiX,德国)评估临床上的屈光不正。>方法:将PowerRefractor的有效性和可重复性与以下各项进行比较:主观(非睫状肌麻痹)屈光度为100名成人受试者(平均年龄23.8(SD 5.7)岁)和客观自动屈光度(Shin-Nippon SRW-5000,日本)为150名受试者(20.1(4.2)岁)。通过检查从每只眼睛获取的自动验光仪读数之间的差异,以及在随后的会议中重新测量100眼的客观处方,来评估可重复性。>结果:平均而言,PowerRefractor处方与验光师的处方之间没有显着差异主观验光,尽管变化很大(差异+0.05(0.63)D,p = 0.41),并且比SRW-5000处方更负(通过-0.20(0.72)D,p <0.001)。关于屈光不正的类型或大小,仪器的精度没有明显的偏差。发现PowerRefractor在所检查的−8.75D至+ 4.00D(平均球当量)的处方范围内具有可重复性。>结论: PowerRefractor是一种有用的物镜检查工具,因为它具有远程和快速功能同时测量双眼能够评估在各种不受限制的观察条件和患者类型下的动眼反应。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号