The International Atomic Energy Agency (IAEA) has organized an international comparison to assess Single Photon Emission Computed Tomography (SPECT) image quantification capabilities in 12 countries. Iodine-131 was chosen as the radionuclide for the comparison because of its wide use around the world, but for logistical reasons solid 133Ba sources were used as a long-lived surrogate for 131I. For this study, we designed a set of solid cylindrical sources so that each site could have a set of phantoms (having nominal volumes of 2 mL, 4 mL, 6 mL, and 23 mL) with traceable activity calibrations so that the results could be properly compared. We also developed a technique using two different detection methods for individually calibrating the sources for 133Ba activity based on a National standard. This methodology allows for the activity calibration of each 133Ba source with a standard uncertainty on the activity of 1.4 % for the high-level 2-, 4-, and 6-mL sources and 1.7 % for the lower-level 23 mL cylinders. This level of uncertainty allows for these sources to be used for the intended comparison exercise, as well as in other SPECT image quantification studies.
机译:国际原子能机构(IAEA)已组织了一次国际比较,以评估12个国家的单光子发射计算机断层扫描(SPECT)图像量化能力。由于碘131在世界范围内的广泛使用,因此选择碘131作为比较的放射性核素,但出于后勤原因,固体 133 Ba源被用作 131 的长寿命替代物。 sup>我在本研究中,我们设计了一组实心圆柱体光源,以便每个位点都可以具有一组幻像(标称体积分别为2 mL,4 mL,6 mL和23 mL),并具有可追溯的活性标定值,因此结果可以是适当比较。我们还开发了一种使用两种不同检测方法的技术,用于根据国家标准分别校准 133 Ba活性的来源。这种方法可以对每个 133 Ba离子源的活性进行校准,标准不确定度是高浓度的2-,4-和6-mL离子源的活性为1.4%,而高浓度的2-mL离子源的活性为1.7%。较低级别的23 mL气瓶。这种不确定性水平允许将这些来源用于预期的比较练习以及其他SPECT图像量化研究中。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号