Comparison of the NIST and NPL Air Kerma Standards Used for X-Ray Measurements Between 10 kV and 80 kV

机译:用于10 kV和80 kV之间的X射线测量的NIST和NPL空气科尔玛标准的比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A direct comparison was made between the air kerma primary standards used for the measurements of low-energy x rays at the National Institute of Standards and Technology (NIST) and the National Physical Laboratory (NPL). The comparison was conducted at the NPL using NPL reference radiation qualities between 10 kV and 80 kV. The results show the primary air-kerma standards to agree within 0.6 % of their values for beam qualities up to 80 kV.
机译:在美国国家标准技术研究院(NIST)和国家物理实验室(NPL)的用于测量低能X射线的空气比释动能基本标准之间进行了直接比较。比较是在NPL中使用10 kV和80 kV之间的NPL参考辐射质量进行的。结果表明,对于高达80 kV的光束质量,主要的空气比释动能标准在其值的0.6%内一致。