Comparison of the NIST and NPL Air kerma standards used for X-Ray Measurements Between 10kV and 80kV

机译:用于10 kV和80 kV之间的X射线测量的NIST和NPL空气比释动能标准的比较

获取原文
获取原文并翻译 | 示例
           

摘要

A direct comparison was made between the air kerma primary standards used for the measurements of low-energy x rays at the National Institute of Standards and Technology (NIST)and the National Physical Laboratory(NPL).The comparison was conducted at the NPL using NPL reference radiation qualities between 10kV and 80kV.The results show the primary airkerma standards to agree within 0.6% of their values for beam qualities up to 80kV.
机译:在美国国家标准技术研究所(NIST)和国家物理实验室(NPL)的用于测量低能X射线的空气比释动能基本标准之间进行了直接比较。在NPL中使用NPL进行了比较参考辐射质量在10kV和80kV之间。结果表明,对于80kV以下的光束质量,主要的航空突释标准在其值的0.6%之内。