首页> 美国卫生研究院文献>Case Reports in Urology >Autoinflation Leading to Failure of Two-Piece Ambicor Implantable Penile Prosthesis: An Outcome from a Methodical Treatment of Recalcitrant Stuttering Priapism
【2h】

Autoinflation Leading to Failure of Two-Piece Ambicor Implantable Penile Prosthesis: An Outcome from a Methodical Treatment of Recalcitrant Stuttering Priapism

机译:自动充气导致两件式Ambicor可植入阴茎假体失败:系统地治疗顽固性口吃性Pri肌的结果

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Introduction. We present the case of a patient who received a two-piece Ambicor penile prosthesis for idiopathic recurrent “stuttering” priapism refractory to other treatment options. The patient returned unable to deflate the device due to an interesting anatomically induced mechanical failure. Aims. To describe the method and findings of this inflatable prosthesis failure. Results. Prosthesis failure occurred due to restrictive corporal diameter and the unique characteristics of fluid reservoir location in the two-piece inflatable prosthesis. The patient was successfully converted to a semirigid prosthesis with resolution of the pain that was due to his prosthesis autoinflation. Conclusion. Stuttering priapism remains a challenging clinical problem. Penile implantation is a reasonable long-term solution in a patient refractory to less invasive options. In patients with fibrotic corpora, a malleable device should be considered (at least temporarily) if unable to dilate comfortably to 13 mm.
机译:介绍。我们介绍了这样一个患者,该患者接受了两件式Ambicor阴茎假体,该假体因特发性复发性“口吃”阴茎异常勃勃而难以治愈。由于有趣的解剖学上引起的机械故障,患者返回无法使装置放气。目的描述这种充气假体失败的方法和发现。结果。假体失败是由于受限的体径和两件式充气假体中储液器位置的独特性所致。该患者成功地转换为半刚性假体,并缓解了由于其假体自体膨胀而引起的疼痛。结论。口吃的狂妄症仍然是一个具有挑战性的临床问题。对于难于侵入性较小的患者,阴茎植入术是合理的长期解决方案。对于患有纤维化体的患者,如果不能舒适地扩张至13mm,应考虑使用可延展的器械(至少是暂时性的)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号