In-Line Measurement of Water Content in Ethanol Using a PVA-Coated Quartz Crystal Microbalance

机译:使用PVA涂层石英晶体微量天平在线测量乙醇中的水分

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

An in-line device for measuring the water content in ethanol was developed using a polyvinyl alcohol (PVA)-coated quartz crystal microbalance. Bio-ethanol is widely used as the replacement of gasoline, and its water content is a key component of its specifications. When the PVA-coated quartz crystal microbalance is contacted with ethanol containing a small amount of water, the water is absorbed into the PVA increasing the load on the microbalance surface to cause a frequency drop. The determination performance of the PVA-coated microbalance is examined by measuring the frequency decreases in ethanol containing 2% to 10% water while the ethanol flows through the measurement device. The measurements indicates that the higher water content is the more the frequency reduction is, though some deviation in the measurements is observed. This indicates that the frequency measurement of an unknown concentration of water in ethanol can be used to determine the water content in ethanol. The PVA coating is examined by microscopy and FTIR (Fourier transform infrared) spectroscopy.
机译:使用涂有聚乙烯醇(PVA)的石英晶体微量天平开发了一种用于测量乙醇中水含量的在线设备。生物乙醇被广泛用作汽油的替代品,其含水量是其规格的关键组成部分。当将涂有PVA的石英微天平与包含少量水的乙醇接触时,水会吸收到PVA中,从而增加了微天平表面的负荷,从而导致频率下降。通过测量当乙醇流过测量设备时含水量为2%到10%的乙醇中的频率降低来检查PVA涂层微量天平的测定性能。测量结果表明,水含量越高,频率降低的幅度越大,尽管在测量结果中会观察到一些偏差。这表明对乙醇中未知水浓度的频率测量可用于确定乙醇中的水含量。通过显微镜和FTIR(傅里叶变换红外)光谱检查PVA涂层。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号