Partial outflow pump dehiscence following off-pump HeartWare left ventricular assist device implantation in a patient with the Marfan syndrome

机译:体外反流HeartWare左心室辅助装置植入术后部分流出泵开裂

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A 30-year-old man with Marfan syndrome, biventricular heart failure, and an earlier left ventricular assist device placement presented to the emergency department with chest discomfort. Computed tomography with contrast showed contrast extravasation from the anterior aspect of the ascending aorta. The patient was immediately taken to the operating room, where re-exploration of the mediastinum disclosed two needle holes from the distal anastomosis at the toe of the graft that were no longer hemostatic, and it appeared that the patient’s heart had dilated in the interim to the point that the outflow graft was too snug. In the operating room, it appeared that a markedly dilated right ventricle was stretching the graft. The patient underwent lengthening of the left ventricular outflow graft and fully recovered.
机译:一名30岁的男子患有Marfan综合征,双心室心力衰竭和较早放置左心室辅助装置,因胸部不适而出现在急诊科。对比计算机断层扫描显示升主动脉的前侧有造影剂外渗。患者立即被送往手术室,在对纵隔进行重新探查时发现,远端吻合处的两个针孔在移植物的脚趾处不再止血,在此期间,患者的心脏似乎已经扩张。流出移植物太贴身的点。在手术室中,似乎明显扩张的右心室正在拉伸移植物。患者接受了延长的左心室流出物移植并完全康复。