Background:Location of the affected bronchus of pleural air leaks is the most important step of trans-bronchoscopic bronchial occlusion for the treatment of intractable pneumothorax. The balloon occlusion test is the most commonly used technique, but has failed in some cases. The aim of the present study was: (1) to determine if endo-bronchial end-tidal CO2 (EtCO2) measurement can identify the affected bronchus that is the source of a persistent pleural air leak; and (2) to establish a methodology for endo-bronchial EtCO2 testing in locating affected bronchus in intractable pneumothorax.
机译:背景:受影响的胸膜漏气支气管的位置是经支气管镜支气管阻塞治疗顽固性气胸的最重要步骤。气囊阻塞测试是最常用的技术,但在某些情况下失败了。本研究的目的是:(1)确定支气管内潮气中二氧化碳(EtCO2)的测量是否可以确定受影响的支气管,这是持续性胸膜漏气的来源; (2)建立用于支气管内EtCO2检测的方法,以在难治性气胸中定位受影响的支气管。