Effect of endoscopic augmentation of the lower oesophageal sphincter (Gatekeeper reflux repair system) on intraoesophageal dynamic characteristics of acid reflux

机译:内镜下食管括约肌增强术(Gatekeeper返流修复系统)对食管内酸返流动态特性的影响

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Background and aims: Improvements in symptoms following endoscopic procedures for gastro-oesophageal reflux disease (GORD) are seldom supported by normalisation of acid exposure time at the distal oesophagus. However, the distribution of gastric acid within the proximal oesophagus is a main determinant of symptom generation in GORD patients. In this study, our aim was to assess the effect of endoscopic insertion of hydrogel expandable prostheses into the oesophageal submucosa on spatiotemporal characteristics of gastro-oesophageal reflux.>Methods: Oesophageal manometry and multichannel ambulatory 24 hour pH monitoring were carried out in nine patients before and six months after the endoscopic procedure. Dynamic characteristics of gastro-oesophageal reflux in patients were also compared with those in 13 asymptomatic controls.>Results: Acid exposure time (AET) at the distal oesophagus decreased from 11.7% (95% confidence interval 6.1–21.8) at baseline to 7.7% (3.7–11.6) at follow up (NS). Of the nine patients, distal AET normalised in three. AET at the middle (7.6% (2.9–12.3)) and proximal (2.4% (0.1–4.8)) oesophagus decreased significantly in all patients (2.4% (0.3–4.5), p <0.01; 1.2% (0.2–2.2), p<0.05 respectively). Proximal extent of acid events significantly decreased in all patients at follow up (37.3% v 9.5%), reaching values observed in asymptomatic controls. Median GORD health related quality of life scores significantly improved from 35.5 at baseline to 9.4.>Conclusions: Despite the lack of a significant improvement in traditional pH variables, endoscopic implant of hydrogel prostheses above the lower oesophageal sphincter significantly decreases proximal spread of acid reflux into oesophageal body. This effect would explain the improvement in symptoms in patients six months after therapy.
机译:>背景和目标:很少通过远端食管的酸暴露时间正常化来支持内镜治疗胃食管反流病(GORD)后症状的改善。然而,胃酸在近端食管中的分布是GORD患者症状发生的主要决定因素。在这项研究中,我们的目的是评估内窥镜插入水凝胶可膨胀假体插入食管粘膜对胃食管反流的时空特征的影响。>方法:食管测压法和多通道动态24小时pH监测在内镜手术之前和之后六个月对9例患者进行了检查。还比较了13位无症状对照患者的胃食管反流的动态特征。>结果:食管远端的酸暴露时间(AET)从11.7%(95%置信区间6.1–21.8)降低了)在基线时为7.7%(3.7–11.6),在随访中(NS)。在9例患者中,有3例远端AET恢复正常。在所有患者中,食道中段(7.6%(2.9–12.3))和近端(2.4%(0.1–4.8))的AET显着降低(2.4%(0.3–4.5),p <0.01; 1.2%(0.2–2.2) ,分别为p <0.05)。在随访中,所有患者的酸事件近端程度均显着降低(37.3%vs 9.5%),达到无症状对照中观察到的值。 GORD健康相关的生活质量中位数从基线的35.5显着提高到9.4。>结论:尽管传统pH值缺乏显着改善,但内窥镜下食管下括约肌上方的水凝胶假体植入明显减少酸反流向近端扩散到食管体内。这种作用可以解释治疗六个月后患者症状的改善。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号