Newborn Screening for Congenital Hypothyroidism-Clinical Evaluation and Comparison of Two Different Test Kits for the Determination of TSH in Dried Blood Samples on Two Different Platforms

机译:先天性甲状腺功能减退症的新生儿筛查 - 两种不同检测试剂盒在两个不同平台上测定干血样本中 TSH 的临床评估和比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Newborn screening (NBS) for congenital hypothyroidism (CH) started in the 1970s, with the introduction of radioimmuno assays (RIA) for the measurement of thyroxine (T4), and thyroid stimulating hormone (TSH). With the development of sensitive enzyme immune assays (EIA, FIA, FEIA), RIAs were replaced in the newborn screening laboratories. With the increasing number of analytes and centralization of NBS, there is a growing demand of total automation. In the course of method validation, two fully automated platforms for the determination of TSH in dried blood samples (DBS) were compared. The GSP from PerkinElmer (PE), and the NS2400 from Labsystems (LDx), together with the recommended test kits from both manufacturers. Both systems showed good performance, with recoveries, of 103.0% (LDx) and 98.5% (PE), and CVs for intra and interassay variations at various concentrations, between 4.3 and 15.7. Both assays had a good correlation (r2 = 0.8814). With LDx/NS2400 platform, TSH values were in the mean 2.09 mU/L higher; however, the difference of both results from the mean was within ±2 SD, up to 30 mU/L, and only for values above 50 mU/L did the difference become bigger. However, this has no influence on the clinical interpretation. No false negative results were observed with either of the two platforms. TSH results obtained with the LDx/NS2400 were slightly higher than those obtained with the PE/GSP; however, the recall rate was lower: 0.059% compared to 0.063%. This can be explained by the much narrower distribution of TSH values. In conclusion, both platforms are equally suitable for medium and large NBS laboratories. However, due to the more open structure the LDx/NS2400 platform has a lot of advantages compared to the totally closed PE/GSP platform.

著录项

  • 期刊名称 International Journal of Neonatal Screening
  • 作者

    Ralph Fingerhut;

  • 作者单位
  • 年(卷),期 2021(7),3
  • 年度 2021
  • 页码 51
  • 总页数 11
  • 原文格式 PDF
  • 正文语种
  • 中图分类
  • 关键词

    机译:先天性甲状腺功能减退症 (CH)、新生儿筛查 (NBS)、促甲状腺激素 (TSH)、干血样本 (DBS);
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号