Exhaled nitric oxide in mylar balloons: influence of storage time humidity and temperature.

机译:聚酯薄膜球囊呼出的一氧化氮:储存时间湿度和温度的影响。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

BACKGROUND: Mylar balloons are used to collect exhaled air for analysis of fractional nitric oxide concentration (FENO). AIM: We studied the effect of storage conditions on the stability of nitric oxide (NO) in mylar balloons. METHODS: Exhaled air samples and calibration gases were stored in mylar balloons at 4, 21 and 37 degrees C, with or without silica gel. NO was measured after 0, 6, 9, 24 and 48 h. Scheffe F-tests were used to compare NO values. RESULTS: NO remained stable in balloons for 9 h at all temperatures, without silica gel. NO increased between 9 and 48 h, but only with low initial FENO. Silica gel increased variability. CONCLUSIONS: FENO in mylar balloons is stable for at least 9 h. The storage temperature is not critical, but silica gel increases variability.
机译:背景:聚酯薄膜球囊用于收集呼出的空气,以分析一氧化氮浓度(FENO)。目的:我们研究了储存条件对聚酯薄膜球囊中一氧化氮(NO)稳定性的影响。方法:将呼出的空气样本和校准气体分别在4、21和37摄氏度下(无论是否装有硅胶)存储在聚酯薄膜球囊中。在0、6、9、24和48小时后测量NO。 Scheffe F检验用于比较NO值。结果:在没有硅胶的情况下,NO在气球中在所有温度下均保持稳定9小时。 NO在9至48小时之间增加,但仅在初始FENO较低的情况下增加。硅胶增加了变异性。结论:聚酯薄膜球囊中的FENO至少可稳定9 h。储存温度并不严格,但是硅胶会增加可变性。